Department of Pneumology and Critical Care Medicine, Helios Klinik Ambrock, Chair of Internal Medicine I, Witten/Herdecke University, Ambrocker Weg 60, 58091, Hagen, Germany,
Lung. 2013 Dec;191(6):611-7. doi: 10.1007/s00408-013-9502-4. Epub 2013 Aug 30.
Endobronchial ultrasound-guided transbronchial needle aspiration (EBUS-TBNA) is a standard procedure for intrathoracic lymph node biopsies. The newly developed cryo-needle operates in a similar way to the EBUS-TBNA but is able to obtain specimens for histological evaluation. The purpose of this animal study was to evaluate the feasibility, effect, and safety of the cryo-needle biopsies.
Four EBUS-guided cryo-needle biopsies were obtained from a mediastinal lymph node of a healthy pig. In an open surgery approach, cryo-needle biopsies using activation times of 1, 2, and 3 s (A1/A2/A3) and needle biopsies using a 21-gauge EBUS-TBNA needle were obtained from mesenteric lymph nodes. Cryo-needle biopsies A2 were performed with (A2+) and without (A2-) an oversheath. The size, weight, percentage of lymphatic tissue and artefact-free area of each cryobiopsy were evaluated. Smears were made with the TBNA-needle aspirates to determine the number of lymphocytes per high-power field (HPF). The bleeding duration was measured.
We successfully obtained EBUS-guided cryo-needle biopsies. The area and weight of the biopsies A3 and A2+ were significantly larger compared with A1 (1.7 ± 0.8 and 1.4 ± 0.3 vs. 0.9 ± 0.4 mm(2); 5.2 ± 2.4 and 3.4 ± 1.8 vs. 1.5 ± 0.7 mg). The percentage of lymphatic tissue of the cryobiopsies was 90 ± 25 and 98 % of samples were artefact-free. The number of lymphocytes/HPF of TBNA-needle smears was 128 ± 54.3. There was no difference in bleeding duration between the techniques.
The cryo-needle yields large histological specimens of high quality.
经支气管超声引导针吸活检术(EBUS-TBNA)是一种用于胸内淋巴结活检的标准程序。新开发的冷冻活检针的操作方式与 EBUS-TBNA 相似,但能够获得用于组织学评估的标本。本动物研究的目的是评估冷冻活检针的可行性、效果和安全性。
对一只健康猪的纵隔淋巴结进行了 4 次 EBUS 引导下的冷冻针活检。在开放式手术中,使用 1、2 和 3 秒激活时间(A1/A2/A3)从肠系膜淋巴结获得冷冻针活检,并用 21 号 EBUS-TBNA 针进行针活检。进行了冷冻针活检 A2 并带有(A2+)和不带有(A2-)护套。评估每个冷冻活检的大小、重量、淋巴组织百分比和无伪影区域。用 TBNA 针抽吸物制作涂片,以确定每高倍镜视野(HPF)的淋巴细胞数。测量出血时间。
我们成功地获得了 EBUS 引导下的冷冻针活检。活检 A3 和 A2+的面积和重量明显大于 A1(1.7±0.8 和 1.4±0.3 与 0.9±0.4mm(2);5.2±2.4 和 3.4±1.8 与 1.5±0.7mg)。冷冻活检的淋巴组织百分比为 90±25%,98%的样本无伪影。TBNA 针涂片的淋巴细胞/HPF 数为 128±54.3。两种技术的出血时间无差异。
冷冻活检针可获得高质量的大组织标本。