Randall A G, Garcia-Webb P, Beilby J P
Department of Clinical Biochemistry, Queen Elizabeth II Medical Centre, Perth, Western Australia.
Ann Clin Biochem. 1990 Jul;27 ( Pt 4):345-52. doi: 10.1177/000456329002700411.
As part of an evaluation of a Synchron CX5 analyser (Beckman Instruments Inc, Brea, USA) we examined a range of tests for interference from haemolysis, bilirubin and lipaemia. Tests investigated were urea, creatinine, urate, total protein, albumin, calcium, total bilirubin, alkaline phosphatase (ALP), aspartate transaminase (AST), gamma-glutamyl transferase (GGT) and inorganic phosphate. Two types of interferences were found. One type is found on other analysers and represents analytical difficulties with the measurement of that particular analyte. The other type of interference was a consequence of the bichromatic optical system used on the CX-5. This latter group includes haemoglobin interference in the measurement of total protein and inorganic phosphate, and bilirubin interference with the measurement of total protein, glucose and inorganic phosphate. Lipaemia interfered with total protein, total bilirubin, inorganic phosphate, urate and glucose. Alternative and modified methods are proposed to improve the measurement of total protein, glucose, total bilirubin and inorganic phosphate. The use of the modified methods for glucose, inorganic phosphate and total bilirubin are limited, at this time, by an error in the calculation algorithm used by the analyser for two step or triggered chemistries, and to a lesser extent, by a reduction in sample throughput.
作为对Synchron CX5分析仪(美国加利福尼亚州布雷亚市贝克曼仪器公司)评估的一部分,我们检测了一系列受溶血、胆红素和脂血干扰的测试项目。所检测的项目包括尿素、肌酐、尿酸、总蛋白、白蛋白、钙、总胆红素、碱性磷酸酶(ALP)、天冬氨酸转氨酶(AST)、γ-谷氨酰转移酶(GGT)和无机磷酸盐。发现了两种干扰类型。一种干扰类型在其他分析仪上也存在,代表了对该特定分析物测量的分析困难。另一种干扰类型是CX - 5所使用的双色光学系统导致的。后一组干扰包括血红蛋白对总蛋白和无机磷酸盐测量的干扰,以及胆红素对总蛋白、葡萄糖和无机磷酸盐测量的干扰。脂血干扰了总蛋白、总胆红素、无机磷酸盐、尿酸和葡萄糖。我们提出了替代方法和改进方法以改善总蛋白、葡萄糖、总胆红素和无机磷酸盐的测量。目前,由于分析仪用于两步或触发式化学分析的计算算法存在误差,且在较小程度上由于样本通量降低,葡萄糖、无机磷酸盐和总胆红素改进方法的使用受到限制。