Clinical Microbiology Service, Department of Laboratory Medicine, Memorial Sloan-Kettering Cancer Center, New York, New York, USA.
J Clin Microbiol. 2014 Jan;52(1):288-90. doi: 10.1128/JCM.02787-13. Epub 2013 Oct 16.
We evaluated the performance characteristics of the FilmArray respiratory panel and the eSensor respiratory viral panel on clinical and spiked lower respiratory tract specimens (LRTS). The overall agreement between the two methods was 89.5% (51/57). The lower limit of detection of both assays for all targets in LRTS was comparable to that for nasopharyngeal swab specimens.
我们评估了 FilmArray 呼吸道 panel 和 eSensor 呼吸道病毒 panel 在临床和加标下呼吸道标本(LRTS)中的性能特征。两种方法的总体一致性为 89.5%(51/57)。两种检测方法在 LRTS 中所有靶标的检测下限与鼻咽拭子标本相当。