石英-钨-卤素聚合装置中的光强衰减

Light intensity decay in quartz-tungsten-halogen polymerization units.

作者信息

H Al-Samadani Khalid, Ayman Al-Dharrab, Mohammed A Wahbi, Loay I Algizani

机构信息

Department of Restorative Dental Science, College of Dentistry, Taibah University, Al Madinah Al Munawwarah - 43353, Kingdom of Saudi Arabia.

出版信息

J Int Oral Health. 2013 Feb;5(1):23-30. Epub 2013 Feb 26.

Abstract

OBJECTIVES

The aim of this study was to determine the effect of ageing on the light intensity of quartz-tungsten-halogen (QTH) polymerization units in dental schools and dental clinics.

MATERIALS & METHODS: Two dental schools and multiple public health dental clinics were visited and the light intensity of curing units was assessed. The age, make, model and service history were recorded. A calibrated radiometer (range 0-1000 nm) was used to measure the light intensity of 171 QTH units. At the end of each visit, the team members provided each dental clinic with a verbal report of the test results and recommendations on maintenance of light polymerization units.

RESULTS

The 171 units tested were from seven different manufacturers. The mean light intensity was 373.1 mW/cm(2), the lowest intensity being 95.3 and the highest 526.60 mW/cm(2). Only 29 QTH units recorded more than 500 mW/cm(2); 102 units showed readings that ranged between 300 and 500 mW/cm(2) and 39 units recorded readings below 300 mW/cm(2). It was not able that one unit had light intensity reading of less than 100 mW/cm(2). The age distribution was as follows: 35 units were more than 10 years old, 79 were from 5 to 10 years old and 57 were less than 5 years old. There was a high statistical significant relation between the mean light intensity and age (P = 0.001).

CONCLUSION

Light intensity values used in some dental clinics were to be found lower than what is recommended for optimum polymerization of resin-based composite restorations. Awareness among dentists of the value of QTH light curing units and their commitment to care and maintenance are crucial to achieving optimum restoration properties. How to cite this article: Al-Samadani KH, Al-Dharrab A, Wahbi MA , Algizani L. Light Intensity Decay in Quartz-Tungsten-Halogen Polymerization Units. J Int Oral Health 2013; 5(1):23-30.

摘要

目的

本研究旨在确定老化对牙科学院和牙科诊所中石英钨卤素(QTH)聚合单元光强度的影响。

材料与方法

走访了两所牙科学院和多家公共卫生牙科诊所,并评估了固化单元的光强度。记录了其使用年限、品牌、型号和维修历史。使用校准过的辐射计(范围为0 - 1000纳米)测量了171个QTH单元的光强度。每次走访结束时,团队成员向各牙科诊所提供了测试结果的口头报告以及关于光聚合单元维护的建议。

结果

所测试的171个单元来自七个不同的制造商。平均光强度为373.1毫瓦/平方厘米,最低强度为95.3,最高为526.60毫瓦/平方厘米。只有29个QTH单元的记录超过500毫瓦/平方厘米;102个单元的读数在300至500毫瓦/平方厘米之间,39个单元的读数低于300毫瓦/平方厘米。没有一个单元的光强度读数低于100毫瓦/平方厘米。使用年限分布如下:35个单元使用超过10年,79个单元使用5至10年,57个单元使用不到5年。平均光强度与使用年限之间存在高度统计学显著关系(P = 0.001)。

结论

发现一些牙科诊所使用的光强度值低于树脂基复合修复体最佳聚合所需的推荐值。牙医对QTH光固化单元价值的认识以及他们对其保养维护的重视对于获得最佳修复性能至关重要。如何引用本文:Al-Samadani KH, Al-Dharrab A, Wahbi MA, Algizani L. 石英钨卤素聚合单元中的光强度衰减。《国际口腔健康杂志》2013年;5(1):23 - 30。

相似文献

10
Polymerization efficiency of LED curing lights.发光二极管固化灯的聚合效率。
J Esthet Restor Dent. 2002;14(5):286-95. doi: 10.1111/j.1708-8240.2002.tb00524.x.

本文引用的文献

10
Contemporary issues in photocuring.
Compend Contin Educ Dent Suppl. 1999(25):S4-15; quiz S73.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验