Suppr超能文献

圆锥角膜眼中使用旋转式Scheimpflug相机、镜面显微镜、光学低相干反射测量法和超声测厚法测量中央角膜厚度的比较。

Comparison of central corneal thickness measurements with a rotating scheimpflug camera, a specular microscope, optical low-coherence reflectometry, and ultrasound pachymetry in keratoconic eyes.

作者信息

Cinar Yasin, Cingu Abdullah Kursat, Turkcu Fatih Mehmet, Cinar Tuba, Sahin Alparslan, Yuksel Harun, Ari Seyhmus

机构信息

Department of Ophthalmology, Faculty of Medicine, Dicle University , Diyarbakir , Turkey and.

出版信息

Semin Ophthalmol. 2015 Mar;30(2):105-11. doi: 10.3109/08820538.2013.833264. Epub 2013 Oct 30.

Abstract

OBJECTIVE

To compare central corneal thickness (CCT) measurements with a rotating Scheimpflug camera (RSC), noncontact specular microscopy (SM), optical low-coherence reflectometry (OLCR), and ultrasonic pachymetry (UP) in keratoconus (KC) patients.

METHOD

In this prospective study, four CCT measurements taken with an RSC, SM, OLCR, and UP were compared in 81 eyes of 44 consecutive KC patients. The KC patients were divided into four subgroups according to Amsler-Krumeich's KC classification.

RESULTS

The RSC and UP measurements of the CCT were not statistically significant in all the groups. Comparison of the SM vs. the OLCR measurements yielded statistically significant differences in all the KC patients and in all KC stages. In all the KC patients, RSC and OLCR showed a high correlation coefficient factor (r = 0.87, p = 0.000).

CONCLUSION

CCT measurements with RSC are comparable to those achieved with UP. Compared with the other devices, according to SM measurements, the central cornea is thicker in all keratoconic eyes and in all KC grades, and it is thinner according to OLCR. RSC, UP, SM, and OLCR should not be used interchangeably in keratoconic eyes.

摘要

目的

比较旋转式Scheimpflug相机(RSC)、非接触式角膜内皮显微镜(SM)、光学低相干反射测量法(OLCR)和超声角膜测厚法(UP)在圆锥角膜(KC)患者中测量中央角膜厚度(CCT)的结果。

方法

在这项前瞻性研究中,对44例连续KC患者的81只眼分别用RSC、SM、OLCR和UP进行了4次CCT测量。根据Amsler-Krumeich的KC分类法,将KC患者分为四个亚组。

结果

所有组中,RSC和UP测量的CCT无统计学差异。在所有KC患者及所有KC分期中,SM与OLCR测量结果比较有统计学差异。在所有KC患者中,RSC和OLCR显示出较高的相关系数因子(r = 0.87,p = 0.000)。

结论

RSC测量的CCT与UP测量结果相当。与其他设备相比,根据SM测量结果,所有圆锥角膜眼及所有KC分级的中央角膜均更厚,而根据OLCR测量则更薄。在圆锥角膜眼中,RSC、UP、SM和OLCR不应互换使用。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验