Suppr超能文献

使用旋转式Scheimpflug相机和超声测厚法测量圆锥角膜中央角膜厚度的可重复性和重复性

Reproducibility and repeatability of central corneal thickness measurement in keratoconus using the rotating Scheimpflug camera and ultrasound pachymetry.

作者信息

de Sanctis Ugo, Missolungi Alessandro, Mutani Bernardo, Richiardi Lorenzo, Grignolo Federico M

机构信息

Department of Clinical Physiopathology, Ophthalmology Institute, University of Turin, Turin, Italy.

出版信息

Am J Ophthalmol. 2007 Nov;144(5):712-718. doi: 10.1016/j.ajo.2007.07.021. Epub 2007 Sep 17.

Abstract

PURPOSE

To assess repeatability, reproducibility, and agreement of rotating Scheimpflug camera (Pentacam Oculus, Wetzlar, Germany) and ultrasound pachymetry in measuring central thickness of keratoconic corneas.

DESIGN

Method-comparison study.

METHODS

In 33 patients with keratoconus (one eye per patient), two examiners each used both pachymetric methods to measure central corneal thickness (CCT); in the same session, measurements then were repeated by examiner 1 (A.M.). The difference between two examiners, and between first and second measurements by examiner 1, with both methods and the difference between the two pachymetric methods in measuring central thickness of keratoconic corneas were noted.

RESULTS

With the rotating Scheimpflug camera, interexaminer correlation was higher (intra-class correlation coefficient [ICC], 0.98 vs 0.76) and inter-examiner variability was lower (95% limits of agreement [95% LoA], -14.8 to 13.8 microm vs -18.0 to +49.5 microm) than with ultrasound pachymetry. Both methods showed close first- to second-measurement correlation (ICC, > 90), but the rotating Scheimpflug camera had lower variability (95% LoA, -14.5 to 14.2 microm vs -27.4 to 26.0 microm). Mean CCT was 478.9 +/- 34.6 microm with the rotating Scheimpflug camera and 486.6 +/- 30 microm with ultrasound pachymetry. Although the mean difference was small (-7.8 microm), the 95% LoA (-43.8 to 28.2 microm) showed that the difference between the two methods can be considerable.

CONCLUSIONS

In keratoconic corneas, the rotating Scheimpflug camera provides measurements of central thickness that are more reproducible and repeatable than those obtained with ultrasound pachymetry. The rotating Scheimpflug camera seems to be suitable for disease staging and follow-up, when corneal thickness measurements may be repeated over time by different examiners.

摘要

目的

评估旋转式Scheimpflug相机(德国韦茨拉尔的Oculus Pentacam)和超声测厚法在测量圆锥角膜中央厚度方面的可重复性、再现性及一致性。

设计

方法比较研究。

方法

对33例圆锥角膜患者(每位患者一只眼),两名检查者分别使用两种测厚方法测量中央角膜厚度(CCT);在同一会诊中,检查者1(A.M.)随后重复测量。记录两名检查者之间、检查者1的第一次和第二次测量之间、两种方法在测量圆锥角膜中央厚度时的差异以及两种测厚方法之间的差异。

结果

与超声测厚法相比,旋转式Scheimpflug相机的检查者间相关性更高(组内相关系数[ICC],0.98对0.76),检查者间变异性更低(95%一致性界限[95% LoA],-14.8至13.8微米对-18.0至+49.5微米)。两种方法的第一次测量与第二次测量相关性均较高(ICC,>90),但旋转式Scheimpflug相机的变异性更低(95% LoA,-14.5至14.2微米对-27.4至26.0微米)。旋转式Scheimpflug相机测得的平均CCT为478.9±34.6微米,超声测厚法测得的平均CCT为486.6±30微米。虽然平均差异较小(-7.8微米),但95% LoA(-43.8至28.2微米)表明两种方法之间的差异可能相当大。

结论

在圆锥角膜中,旋转式Scheimpflug相机提供的中央厚度测量结果比超声测厚法更具可重复性和再现性。当角膜厚度测量可能由不同检查者随时间重复进行时,旋转式Scheimpflug相机似乎适用于疾病分期和随访。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验