Abstract

The effects of magnetic resonance imaging were assessed on four dual chamber and two single chamber pulse generators. The tests were performed with a resistive, water-cooled magnet operating at 0.15 T. The 6.4-MHz radiofrequency (RF) field was operated at a maximum power of 1,000 watts with a period adjusted from 130 to 500 ms. Reed switch closure occurred in all six pulse generators tested when placed near the entrance of the magnetic resonance imaging scanner, and the generators reverted to asynchronous operation unless programmed to the "magnet off" mode. None of the pulse generators exhibited any alterations in programmed parameters or in the ability to be reprogrammed after RF pulsing. When the RF field was turned on, there was no change in the asynchronous paced cycle length in four pulse generators; however, during RF scanning there was rapid cardiac stimulation at the RF pulse period in one single chamber and one dual chamber pulse generator.

摘要

对四个双腔和两个单腔脉冲发生器进行了磁共振成像效应评估。测试使用的是一台电阻式水冷磁体,磁场强度为0.15T。6.4MHz射频(RF)场的最大功率为1000瓦,周期从130毫秒调整到500毫秒。当所有六个测试的脉冲发生器放置在磁共振成像扫描仪入口附近时,簧片开关闭合,除非设置为“磁体关闭”模式,否则发生器会恢复到异步运行状态。射频脉冲后,没有一个脉冲发生器的编程参数或重新编程能力出现任何改变。当射频场开启时,四个脉冲发生器的异步起搏周期长度没有变化;然而,在射频扫描期间,一个单腔和一个双腔脉冲发生器在射频脉冲周期出现了快速心脏刺激。