320 排 CT 扫描仪的剂量学评估。

Dosimetric evaluation of a 320 detector row CT scanner unit.

机构信息

S.C. di Fisica Sanitaria, A.O.U. "Ospedali Riuniti" di Trieste, Trieste, Italy.

出版信息

Radiol Oncol. 2011 Mar;45(1):64-7. doi: 10.2478/v10019-010-0049-1. Epub 2010 Nov 25.

Abstract

BACKGROUND

The technologic improvements in Multislice scanners include the increment in the X-ray beam width. Some new CT scanners are equipped with a 320 detector row which allows a longitudinal coverage of 160 mm and a total of 640 slices for a single rotation. When such parameters are used the length of the traditional pencil chamber (10 cm) is no more appropriate to measure the standard weighted computed tomography dose index (CTDI(w)) value. MATERIALS AND METHODS.: Dosimetric measurements were performed on a 640 slices Toshiba Aquilion One CT scanner using common instrumentation available in Medical Physics Departments.

RESULTS

For the measurements in air, two different ionization chambers were completely exposed to the beam. Dosimeters showed an acceptable agreement in the measurements. To evaluate the actual shape of the dose profile strips of Gafchromic XRQA film were used. Films were previously calibrated on site. From the graphic response of the scanned film it is possible to evaluate the full width at half maximum (FWHM) of the dose profile which represent the actual beam width.

CONCLUSIONS

Computed Tomography Dose Index (CTDI) and Dose Length Product (DLP) need to be changed when the beam width of the CT scanner is over 100 mm. To perform dose evaluation with the conventional instrumentation, two parameters should be considered: the average absorbed dose and the actual beam width. To measure the average absorbed dose, the conventional ionization chamber can be used. For the measurement of the width of the dose profile, Gafchromic XRQA film seemed to be suitable.

摘要

背景

多层扫描仪的技术改进包括 X 射线束宽度的增加。一些新型 CT 扫描仪配备了 320 个探测器排,可实现 160mm 的纵向覆盖范围和单次旋转 640 个切片。当使用这些参数时,传统铅笔束腔(10cm)的长度不再适合测量标准加权计算机断层扫描剂量指数(CTDI(w))值。

材料和方法

在东芝 Aquilion One CT 扫描仪上进行了 640 层的剂量学测量,使用医学物理部门常用的仪器。

结果

对于空气测量,两个不同的电离室完全暴露在射束下。剂量计在测量中表现出良好的一致性。为了评估剂量分布带的实际形状,使用了 Gafchromic XRQA 胶片。胶片在现场进行了校准。从扫描胶片的图形响应中,可以评估剂量分布的半最大值全宽(FWHM),它代表实际束宽。

结论

当 CT 扫描仪的束宽超过 100mm 时,需要更改计算机断层扫描剂量指数(CTDI)和剂量长度乘积(DLP)。为了使用常规仪器进行剂量评估,应考虑两个参数:平均吸收剂量和实际束宽。为了测量平均吸收剂量,可以使用常规电离室。对于剂量分布宽度的测量,Gafchromic XRQA 胶片似乎是合适的。