Dilated cardiomyopathy is the most common type of the heart failure which can be characterized by impaired ventricular contractility. Mechanical circulatory support devices were introduced into practice for the heart failure patients to bridge the time between the decision to transplant and the actual transplantation which is not sufficient due to the state of donor organ supply. In this study, the hemodynamic response of a cardiovascular system that includes a dilated cardiomyopathic heart under support of a newly developed continuous flow left ventricular assist device--Heart Turcica Axial--was evaluated employing computer simulations. For the evaluation, a numerical model which describes the pressure-flow rate relations of Heart Turcica Axial, a cardiovascular system model describing the healthy and pathological hemodynamics, and a baroreflex model regulating the heart rate were used. Heart Turcica Axial was operated between 8000 rpm and 11,000 rpm speeds with 1000 rpm increments for assessing the pump performance and response of the cardiovascular system. The results also give an insight about the range of the possible operating speeds of Heart Turcica Axial in a clinical application. Based on the findings, operating speed of Heart Turcica Axial should be between 10,000 rpm and 11,000 rpm.
扩张型心肌病是心力衰竭最常见的类型,其特征为心室收缩功能障碍。机械循环支持装置被引入心力衰竭患者的临床实践中,以在决定移植和实际移植之间的时间间隔内提供支持,由于供体器官供应不足,这段时间往往不够。在这项研究中,采用计算机模拟评估了一种新开发的连续流左心室辅助装置——Heart Turcica Axial 支持下的扩张型心肌病心血管系统的血液动力学反应。为了进行评估,使用了描述 Heart Turcica Axial 的压力-流量关系的数值模型、描述健康和病理血液动力学的心血管系统模型以及调节心率的压力反射模型。Heart Turcica Axial 的转速在 8000 rpm 和 11000 rpm 之间,以 1000 rpm 的增量进行操作,以评估泵的性能和心血管系统的反应。研究结果还提供了关于 Heart Turcica Axial 在临床应用中可能的操作速度范围的见解。基于研究结果,Heart Turcica Axial 的操作速度应在 10000 rpm 和 11000 rpm 之间。