Comparison of central corneal thickness measured by hand-held and desk-mounted ultrasound pachymeters in glaucoma patients.

作者信息

Salim Sarwat, Du Haiming, Wan Jim

机构信息

University of Tennessee Health Science Center , Memphis, TN , USA.

出版信息

Semin Ophthalmol. 2015 Jul;30(4):268-71. doi: 10.3109/08820538.2013.839816. Epub 2014 Jan 10.

Abstract

PURPOSE

The purpose of this study was to compare the accuracy and reproducibility of central corneal thickness measured by hand-held and desk-mounted ultrasound pachymeter in glaucoma patients under treatment.

METHODS

Prospective study of 65 glaucoma patients. Central corneal thickness was measured by two ultrasound pachymeters: the hand-held, portable PachPen (Accutome, Lynwood, WA), and the desk-mounted PacScan 300 (Sonomed, Lake Success, NY).

RESULTS

The mean ± SD central corneal thickness was 526.5 ± 44.8 µm and 530.0 ± 44.7 µm for the hand-held and desk-mounted pachymeters, respectively (p = 0.15). Linear regression analysis revealed a slope of 0.97 with Pearson correlation coefficient of 0.96. Bland-Altman analysis showed a mean difference of measurements by both pachymeters of 3.22 µm with two standard deviations = 9.51. Both instruments showed high intraobserver correlation: 0.972 for the hand-held pachymeter and 0.993 for the desk-mounted pachymeter.

CONCLUSION

Central corneal thickness measurements were comparable with the use of hand-held and desk-mounted ultrasound units in glaucoma patients with good intraobserver reproducibility.