Comparison of optical low coherence reflectometry and ultrasound pachymetry in the measurement of central corneal thickness before and after photorefractive keratectomy.

作者信息

Spadea Leopoldo, Giammaria Daniele, Di Genova Lucia, Fiasca Arianna

机构信息

University of L'Aquila, L'Aquila, Italy.

出版信息

J Refract Surg. 2007 Sep;23(7):661-6. doi: 10.3928/1081-597X-20070901-04.

Abstract

PURPOSE

To compare an ultrasound pachymeter with an optical low coherence reflectometry (OLCR) pachymeter for measuring pre- and postoperative central corneal thickness of patients undergoing photorefractive keratectomy (PRK).

METHODS

In a prospective, noncomparative, interventional study, 48 myopic eyes (mean manifest refractive spherical equivalent: -4.93 +/- 2.93 diopters [D]) of 30 healthy patients underwent PRK. Pre- and postoperative central corneal thickness was measured by two examiners (E1, E2) with an ultrasound pachymeter and an OLCR pachymeter. Agreement and inter-rater repeatability were determined using the comparison method described by Bland and Altman.

RESULTS

The limits of agreement between the two devices ranged from 17.8 microm (E1) to 20.5 microm (E2) preoperatively and from 22.4 microm (E1) to 16.9 microm (E2) postoperatively. The coefficient of inter-rater repeatability ranged from 9.1 microm (ultrasound pachymeter) to 5.4 microm (OLCR pachymeter) preoperatively and from 7.1 microm (ultrasound pachymeter) to 4.7 microm (OLCR pachymeter) postoperatively.

CONCLUSIONS

The OLCR pachymeter seems to show better pre- and postoperative repeatability compared to the ultrasound pachymeter. The agreement between the two devices should be considered acceptable for clinical practice. Photorefractive keratectomy did not affect the postoperative agreement and repeatability of the pachymeters.

摘要

目的

比较超声角膜测厚仪与光学低相干反射测量仪(OLCR)在测量接受准分子激光角膜切削术(PRK)患者术前和术后中央角膜厚度方面的差异。

方法

在一项前瞻性、非对照、干预性研究中,30例健康患者的48只近视眼(平均明显屈光球镜等效值:-4.93±2.93屈光度[D])接受了PRK手术。术前和术后中央角膜厚度由两名检查者(E1、E2)分别使用超声角膜测厚仪和OLCR角膜测厚仪进行测量。使用Bland和Altman描述的比较方法确定一致性和检查者间重复性。

结果

术前,两种仪器之间的一致性界限范围为17.8微米(E1)至20.5微米(E2);术后为22.4微米(E1)至16.9微米(E2)。术前,检查者间重复性系数范围为9.1微米(超声角膜测厚仪)至5.4微米(OLCR角膜测厚仪);术后为7.1微米(超声角膜测厚仪)至4.7微米(OLCR角膜测厚仪)。

结论

与超声角膜测厚仪相比,OLCR角膜测厚仪在术前和术后似乎具有更好的重复性。两种仪器之间的一致性在临床实践中应被认为是可接受的。准分子激光角膜切削术并未影响角膜测厚仪术后的一致性和重复性。