Department of Ophthalmology, La Inmaculada Hospital, Huercal Overa, Almeria, Spain.
Cornea. 2013 Jan;32(1):9-13. doi: 10.1097/ICO.0b013e318242fd0f.
To compare central corneal thickness (CCT) results measured by Fourier-domain optical coherence tomography (FD-OCT) and ultrasound pachymetry (USP) in glaucomatous eyes.
In this prospective, observational cross-sectional study, 80 eyes of 80 patients with primary open-angle glaucoma (POAG) and no other ocular abnormality were selected: 28 were treated with 1 drug (subgroup 1), 32 with 2 drugs (subgroup 2), and 20 with 3 drugs (subgroup 3). CCT was measured by FD-OCT (RTVue OCT) and USP (Pachymeter Reichert IOPac). Bland-Altman plots were used to assess the agreement between both instruments. The differences between CCTs measured by USP and FD-OCT were compared among the 3 subgroups.
The mean CCT was 537.76 ± 32.24 μm and 520.53 ± 30.44 μm for USP and FD-OCT, respectively. A significant difference was found between the mean values obtained by FD-OCT and USP (17.22 ± 7.96 μm, P < 0.001, paired Student t test). A high correlation was obtained for CCT measured by both methods (Pearson correlation coefficient = 0.969; P < 0.001), and there was good agreement between the 2 pachymetry methods. Similar differences in CCT using USP and FD-OCT were found among the 3 treatment subgroups (P > 0.05 in all pairwise comparisons, analysis of variance).
FD-OCT underestimates CCT compared with CCT measured by USP in POAG. Although highly correlated, the difference between these 2 devices can be clinically significant in the context of refractive surgeries in POAG patients but not in intraocular pressure estimation. This difference also seems to be independent of the number of antiglaucoma treatments used.
比较频域光学相干断层扫描(FD-OCT)和超声角膜测厚仪(USP)测量青光眼患者中央角膜厚度(CCT)的结果。
在这项前瞻性、观察性的横断面研究中,选择了 80 只眼的 80 例原发性开角型青光眼(POAG)患者,这些患者均无其他眼部异常:28 例患者接受 1 种药物治疗(亚组 1),32 例患者接受 2 种药物治疗(亚组 2),20 例患者接受 3 种药物治疗(亚组 3)。使用 FD-OCT(RTVue OCT)和 USP(Pachymeter Reichert IOPac)测量 CCT。Bland-Altman 图用于评估两种仪器之间的一致性。比较 3 个亚组中 USP 和 FD-OCT 测量的 CCT 之间的差异。
USP 和 FD-OCT 测量的 CCT 平均值分别为 537.76±32.24μm 和 520.53±30.44μm。FD-OCT 和 USP 获得的平均值之间存在显著差异(17.22±7.96μm,P<0.001,配对学生 t 检验)。两种方法测量的 CCT 相关性较高(Pearson 相关系数=0.969;P<0.001),两种角膜测厚仪之间具有良好的一致性。在 3 个治疗亚组中,USP 和 FD-OCT 测量的 CCT 差异相似(所有两两比较,方差分析 P>0.05)。
与 USP 测量的 CCT 相比,POAG 中 FD-OCT 低估了 CCT。尽管两种设备高度相关,但在 POAG 患者的屈光手术背景下,这两种设备之间的差异可能具有临床意义,但在估计眼内压方面没有意义。这种差异似乎也与使用的抗青光眼治疗数量无关。