Biosens Bioelectron. 2014 May 15;55:301-6. doi: 10.1016/j.bios.2013.12.030. Epub 2013 Dec 21.
This paper describes the development of an immunosensor for detection of anti-hemagglutinin antibodies. Its preparation consists of successive modification steps of glassy carbon electrodes: (i) creation of COOH groups, (ii) covalent immobilization of protein A with EDC/NHS coupling reaction, (iii) covering with anti-His IgG monoclonal antibody, (iv) immobilization of the recombinant His-tagged hemagglutinin (His6-H5 HA), (v) filling free space with BSA. The interactions between two variants of recombinant HA (short and long) from highly pathogenic avian influenza virus H5N1 and the anti-H5 HA monoclonal antibody (Mab 6-9-1) have been explored with electrochemical impedance spectroscopy (EIS). The impedimetric immunosensor displayed a very good detection limit (LOD) of 2.1 pg/mL, the quantification limit (LOQ) of 6.3 pg/mL and a dynamic range from 4 pg/mL to 20 pg/mL. In addition, this analytical device was applied for detection of antibodies against His6-H5 HA in serum of vaccinated hen using serial 10-fold dilutions of serum. The immunosensor proposed was able to detect antibody in hen serum diluted up to 7 × 10(7)-fold. The sensitivity of immunosensor was about four orders of magnitude much better than ELISA.
本文介绍了一种用于检测抗血凝素抗体的免疫传感器的研制。其制备过程包括玻碳电极的连续修饰步骤:(i)COOH 基团的生成,(ii)EDC/NHS 偶联反应使蛋白 A 共价固定,(iii)抗 His6 IgG 单克隆抗体覆盖,(iv)重组 His 标记的血凝素(His6-H5 HA)的固定,(v)BSA 填充自由空间。采用电化学阻抗谱(EIS)研究了两种高致病性禽流感病毒 H5N1 的重组 HA(短和长)变体与抗 H5 HA 单克隆抗体(Mab 6-9-1)之间的相互作用。该阻抗免疫传感器具有非常低的检测限(LOD)为 2.1 pg/mL,定量限(LOQ)为 6.3 pg/mL,动态范围为 4 pg/mL 至 20 pg/mL。此外,该分析装置还用于检测接种母鸡血清中针对 His6-H5 HA 的抗体,采用血清的连续 10 倍稀释。该免疫传感器能够检测到稀释至 7×10(7)倍的母鸡血清中的抗体。免疫传感器的灵敏度比 ELISA 高约四个数量级。