Performance of an automated six-wavelength photometer (Radiometer OSM3) for routine measurement of hemoglobin derivatives.

作者信息

Zijlstra W G, Buursma A, Zwart A

机构信息

Department of Physiology, University of Groningen, The Netherlands.

出版信息

Clin Chem. 1988 Jan;34(1):149-52.

PMID:2448064
Abstract

A new instrument for spectrophotometric determination of oxygen saturation (SO2) and the fractions of carboxyhemoglobin (FHbCO) and methemoglobin (FHi), the Hemoximeter OSM3 (Radiometer), was tested by comparative measurements with a multiwavelength method. We found that it gave reliable results for freshly drawn samples from patients. When samples with artificially induced high proportions of HbCO or Hi, or both, were introduced, results were still sufficiently accurate. The presence of sulfhemoglobin (SHb) seriously interfered with the measurement of SO2, FHbCO, and FHi, but the instrument reliably indicated its presence. Because SHb is rarely encountered, a reliable warning for its presence is adequate for clinical practice. Using the instrument in its fetal mode, we obtained accurate results for blood samples from newborns.

摘要

采用多波长方法进行对比测量,对一种用于分光光度法测定血氧饱和度(SO2)、碳氧血红蛋白分数(FHbCO)和高铁血红蛋白分数(FHi)的新型仪器——血氧计OSM3(雷度米特公司)进行了测试。我们发现,对于从患者身上刚采集的样本,该仪器给出的结果可靠。当引入人工诱导产生高比例HbCO或Hi或两者皆有的样本时,结果仍足够准确。硫血红蛋白(SHb)的存在严重干扰了SO2、FHbCO和FHi的测量,但该仪器能可靠地指示其存在。由于SHb很少出现,因此其存在的可靠警示对于临床实践而言已足够。在胎儿模式下使用该仪器,我们获得了新生儿血样的准确结果。