Am J Ophthalmol. 2014 Jun;157(6):1315-1321.e1. doi: 10.1016/j.ajo.2014.02.031. Epub 2014 Feb 15.
To evaluate the efficacy of an iPod application developed to rapidly measure approximate visual acuity (VA).
Prospective, cross-sectional study.
One hundred four subjects (104 eyes) with a wide range of VA underwent both Early Treatment Diabetic Retinopathy Study (ETDRS) VA testing and iPod application testing. The application presented 4 randomized ETDRS letters at the Snellen equivalent of 20/40 and 20/200 each. In separate analyses, sensitivity and specificity of the iPod test were determined using the ETDRS VA testing results as the gold standard and defining "failure" to see as identifying 1, 2, 3, or all 4 letters incorrectly out of 4 trials. The minimum number of iPod trials necessary to maintain high accuracy was determined by defining a "failure" as getting the first, first 2, first 3, or all trial letters incorrect. In 13 subjects, the time necessary to perform the testing was determined.
Forty-six subjects had an ETDRS Snellen equivalent VA ≥20/40; 45 had a VA <20/40 and ≥20/200; and 13 had a VA <20/200. The mean ± SD testing time for the 3 groups was 43 ± 13, 60 ± 15, and 72 ± 25 seconds, respectively. The highest sensitivity and specificity combination was observed when a "failure" was defined as getting all letters incorrect for both the 20/40 (sensitivity = 98.3%/specificity = 93.5%) and 20/200 (sensitivity/specificity both 92.3%) cutoffs.
An iPod application requiring about a 1-minute testing time provides an objective, portable, rapid, and low-cost method to determine approximate VA, allowing VA testing to be performed efficiently in large surveys and other settings where approximate VA should be measured.
评估一种用于快速测量近似视力(VA)的 iPod 应用程序的疗效。
前瞻性、横断面研究。
104 名受试者(104 只眼)接受了早期糖尿病视网膜病变研究(ETDRS)VA 测试和 iPod 应用程序测试,VA 范围广泛。该应用程序以 20/40 和 20/200 的 Snellen 等效值呈现 4 个随机 ETDRS 字母。在单独的分析中,使用 ETDRS VA 测试结果作为金标准,将“失败”定义为在 4 次试验中识别出 1、2、3 或全部 4 个字母中的 1、2、3 或全部 4 个字母不正确,确定 iPod 测试的敏感性和特异性。通过将“失败”定义为第一次、前两次、前三次或所有三次试验字母不正确来确定维持高准确性所需的最小 iPod 试验次数。在 13 名受试者中,确定了进行测试所需的时间。
46 名受试者的 ETDRS Snellen 等效 VA ≥20/40;45 名受试者的 VA <20/40 且≥20/200;13 名受试者的 VA <20/200。3 组的平均(±SD)测试时间分别为 43±13、60±15 和 72±25 秒。当“失败”定义为两次 20/40(敏感性=98.3%/特异性=93.5%)和 20/200(敏感性/特异性均为 92.3%)截止值的所有字母都不正确时,观察到最高的敏感性和特异性组合。
需要大约 1 分钟测试时间的 iPod 应用程序提供了一种客观、便携、快速且低成本的方法来确定近似 VA,从而可以在大型调查和其他需要测量近似 VA 的环境中有效地进行 VA 测试。