通过心房起搏器电极进行远场QRS波群感知。I. AAI和VDD/DDD起搏中的机制、后果、鉴别诊断及对策

Far-field QRS complex sensing via the atrial pacemaker lead. I. Mechanism, consequences, differential diagnosis and countermeasures in AAI and VDD/DDD pacing.

作者信息

Brandt J, Fåhraeus T, Schüller H

机构信息

Department of Thoracic Surgery, Lund University Hospital, Sweden.

出版信息

Pacing Clin Electrophysiol. 1988 Oct;11(10):1432-8. doi: 10.1111/j.1540-8159.1988.tb04992.x.

Abstract

Unintended sensing of QRS complexes via atrial pacemaker leads may cause disorders of pacemaker function in AAI, VDD and DDD pacing. The consequences of this phenomenon depend upon the pacing mode and the timing of the inappropriate sensing as related to the technical characteristics of the pulse generator. With AAI pacemakers, "inappropriate pacemaker bradycardia" may be seen or P-wave undersensing may be simulated. With VDD and DDD systems a special kind of pacemaker mediated tachycardia or apparent P-wave undersensing may result. With knowledge of the underlying mechanisms, differential diagnosis is possible. The countermeasures available are discussed.

摘要

通过心房起搏器电极意外感知QRS波群可能会导致AAI、VDD和DDD起搏模式下起搏器功能紊乱。这种现象的后果取决于起搏模式以及与脉冲发生器技术特性相关的不适当感知的时机。对于AAI起搏器,可能会出现“不适当的起搏器心动过缓”,或者可能会模拟P波感知不足。对于VDD和DDD系统,可能会导致一种特殊类型的起搏器介导的心动过速或明显的P波感知不足。了解潜在机制后,就可以进行鉴别诊断。文中还讨论了可用的应对措施。