Kantharia B K, Wilbur S L, Padder F A, Pennington J C, Samuels F L, Movsowitz C M, Hessen S E, Saari C, Kutalek S P
Hahnemann University Hospital, Philadelphia, PA 19102-1192, USA.
J Interv Card Electrophysiol. 2001 Mar;5(1):59-66. doi: 10.1023/a:1009805807837.
The normal functioning of dual chamber pacemaker-cardioverter defibrillator (AV pacer/ICD) may be affected by oversensing of the farfield R wave (FFRW) by the atrial channel. This study aimed to investigate whether placement of the AV pacer/ICD's atrial lead at a lateral (LAT) wall location compared to a medial (MED) location i.e. the appendage of the right atrium, would reduce the amplitude of FFRWs but not the nearfield atrial electrograms (AEGMs) during sinus rhythm (SR) and ventricular fibrillation (VF). In 17 patients, real time electrograms were recorded during SR and induced VF through the atrial lead initially at the MED and subsequently at the LAT location. In 10 patients the electrograms in SR were also recorded on a computerized data acquisition and recording system at different band-pass filter settings. Although FFRWs were recorded both at MED and LAT locations, they were much smaller, 3.5+/-4.1mm during SR and 1.7+/-2.2mm during VF at the LAT location. At 30-500Hz band-pass filter, lower amplitudes of FFRWs 0.14+/-0.09 mV were recorded at the LAT location. The V/A ratios of the amplitudes of FFRWs and AEGMs were smaller at the LAT location during SR and VF. The nearfield AEGMs were of similar amplitudes at the MED and LAT locations. These data indicate that lower amplitudes of FFRWs are recorded by placement of the atrial lead at the lateral wall of the right atrium. Oversensing of FFRWs may be prevented to improve functioning of the AV pacer-ICD.
双腔起搏器 - 心脏转复除颤器(房室起搏器/植入式心律转复除颤器,AV pacer/ICD)的正常功能可能会受到心房通道对远场R波(FFRW)过度感知的影响。本研究旨在调查将AV pacer/ICD的心房导线置于外侧(LAT)壁位置(与内侧(MED)位置即右心耳相比)是否会在窦性心律(SR)和心室颤动(VF)期间降低FFRW的幅度,但不降低近场心房电图(AEGM)的幅度。在17例患者中,在SR期间以及通过心房导线诱发VF时,最初在MED位置,随后在LAT位置记录实时电图。在10例患者中,还在不同带通滤波器设置下,使用计算机化数据采集和记录系统记录SR期间的电图。尽管在MED和LAT位置均记录到了FFRW,但在LAT位置时其幅度要小得多,SR期间为3.5±4.1mm,VF期间为1.7±2.2mm。在30 - 500Hz带通滤波器下,在LAT位置记录到的FFRW幅度更低,为0.14±0.09mV。在SR和VF期间,LAT位置的FFRW与AEGM幅度的V/A比值更小。在MED和LAT位置,近场AEGM的幅度相似。这些数据表明,将心房导线置于右心房外侧壁可记录到较低幅度的FFRW。可防止FFRW的过度感知,以改善AV起搏器 - ICD的功能。