Br J Radiol. 2014 Apr;87(1036):20130613. doi: 10.1259/bjr.20130613.
The aim of this study is to examine the effect of iDose4 hybrid iterative reconstruction algorithm (Philips Healthcare, Cleveland, OH) on radiation dose and image quality in chest–abdomen–pelvis (CAP) CT scanning of adult patients.
CAP CT examinations were performed on 99 patients with the use of the “old standard” protocol performing filtered back projection reconstruction algorithm (FBP protocol) and on 84 patients with the use of iDose4 protocol on a 64-multidetector CT. Patients were subdivided into three weight groups (Group 1, 41–60kg; Group 2, 61–90kg; and Group 3, .90kg). Volume CT dose index and dose length product (DLP) were recorded, while effective dose was calculated from DLP measurements. Objective image noise, signal-to-noise ratio and contrast-to-noise ratio were calculated. Two radiologists reviewed images to evaluate image quality.
Compared with the FBP protocol, there was an overall 46.5% significant decrease in effective dose with the use of iDose4 protocol.
Objective image quality was higher in iDose4 images than in FBP images. Subjective image noise, sharpness, contrast and diagnostic confidence scores tended to be better for iDose4 protocol at the decreased radiation exposure level. Artefacts were minor for both protocols.
Our results suggest that the iterative acquisition protocol provides great potential for reducing radiation exposure and maintaining or improving image quality in CAP CT examinations.
本研究旨在探讨 iDose4 混合迭代重建算法(飞利浦医疗,克利夫兰,俄亥俄州)对成人胸部-腹部-骨盆(CAP)CT 扫描的辐射剂量和图像质量的影响。
99 例 CAP CT 检查采用滤波反投影重建算法(FBP 协议)的“旧标准”方案进行,84 例采用 iDose4 协议在 64 层多排 CT 上进行。患者分为三组体重组(Group 1,41-60kg;Group 2,61-90kg;Group 3,.90kg)。记录容积 CT 剂量指数和剂量长度乘积(DLP),并从 DLP 测量值计算有效剂量。计算客观图像噪声、信噪比和对比噪声比。两名放射科医生回顾图像以评估图像质量。
与 FBP 协议相比,使用 iDose4 协议可使有效剂量总体降低 46.5%。
与 FBP 图像相比,iDose4 图像的客观图像质量更高。在降低辐射暴露水平的情况下,主观图像噪声、锐度、对比度和诊断信心评分倾向于 iDose4 协议更好。两种方案的伪影都较小。
我们的结果表明,迭代采集方案在 CAP CT 检查中具有降低辐射暴露和保持或提高图像质量的巨大潜力。