Physical movement sensitive pacing: comparison of two "activity"-triggered pacing systems.

作者信息

Stangl K, Wirtzfeld A, Lochschmidt O, Basler B, Mittnacht A

机构信息

I. Medical Clinic, Technical University Munich, Germany.

出版信息

Pacing Clin Electrophysiol. 1989 Jan;12(1 Pt 1):102-10. doi: 10.1111/pace.1989.12.p1.102.

Abstract

The Activitrax (A) and the Sensolog (S) pulse generator units were compared during bench tests and in patients both under ergometric conditions and during daily activities to determine the susceptibility of both systems to external noise. During the bench tests both pacemakers were mounted on a swingboard and subjected to controlled vibrational force. The patients were divided into two groups of 10 patients each, one group with A and the other with S units; the discriminating power of workload changes was examined under treadmill exercise. In six patients of each group the rate behavior of S and A during daily activities was recorded by Holter monitoring. During bench tests, S and A displayed a frequency-selective mode of operation between 2-70 Hz with the greatest sensitivity between 10-50 Hz. In patients, S showed a significantly higher discriminating power to changes in workload during treadmill exercise. According to y = 0.27x + 85.8 (r = 0.92) the relationship between the power and the pacing rate of S was linear, whereas with y = 0.04x + 89.8 (r = 0.2) the pacing rate of A was almost independent of the given power. The susceptibility to external noise occurring while riding in an auto, streetcar or train with rate increases between 7 ppm and 12 ppm did not significantly differ in both systems. Lying prone on the pacemakers produced brief (less than 1 min) rate increases of 38 ppm (S) and 18 ppm (A), respectively.

摘要

在台架试验中以及在患者处于测力计条件下和日常活动期间,对Activitrax(A)和Sensolog(S)脉冲发生器单元进行了比较,以确定这两种系统对外部噪声的敏感度。在台架试验中,将两个起搏器安装在一个摆动板上,并施加受控的振动力。患者被分为两组,每组10名患者,一组使用A单元,另一组使用S单元;在跑步机运动下检查工作量变化的辨别能力。每组的6名患者在日常活动期间通过动态心电图监测记录S和A的心率行为。在台架试验中,S和A在2 - 70Hz之间显示出频率选择操作模式,在10 - 50Hz之间敏感度最高。在患者中,S在跑步机运动期间对工作量变化显示出明显更高的辨别能力。根据y = 0.27x + 85.8(r = 0.92),S的功率与起搏频率之间的关系是线性的,而对于A,y = 0.04x + 89.8(r = 0.2),其起搏频率几乎与给定功率无关。在乘坐汽车、有轨电车或火车时出现的外部噪声敏感度,两种系统的心率增加在7ppm至12ppm之间,没有显著差异。俯卧在起搏器上分别使心率短暂(小于1分钟)增加38ppm(S)和18ppm(A)。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验