Two-dimensional high spatial-resolution dosimeter using europium doped potassium chloride: a feasibility study.

作者信息

Li H Harold, Driewer Joseph P, Han Zhaohui, Low Daniel A, Yang Deshan, Xiao Zhiyan

机构信息

Department of Radiation Oncology, Washington University School of Medicine, St Louis, MO, USA. Center for Materials Innovation, Washington University, St Louis, MO, USA.

出版信息

Phys Med Biol. 2014 Apr 21;59(8):1899-909. doi: 10.1088/0031-9155/59/8/1899. Epub 2014 Mar 20.

Abstract

Recent research has shown that KCl:Eu²⁺ has great potential for use in megavoltage radiation therapy dosimetry because this material exhibits excellent storage performance and is reusable due to strong radiation hardness. This work reports the authors' attempts to fabricate 2D KCl:Eu²⁺ storage phosphor films (SPFs) using both a physical vapor deposition (PVD) method and a tape casting method. X-ray diffraction analysis showed that a 10 µm thick PVD sample was composed of highly crystalline KCl. No additional phases were observed, suggesting that the europium activator had been completely incorporated into the KCl matrix. Photostimulated luminescence and photoluminescence spectra suggested that F (Cl(-)) centers were the electron storage centers post x-ray irradiation and that Eu²⁺ cations acted as luminescence centers in the photostimulation process. The 150 µm thick casted KCl:Eu²⁺ SPF showed sub-millimeter spatial-resolution. Monte Carlo simulations further demonstrated that the admixture of 20% KCl:Eu²⁺ and 80% low Z polymer binder exhibited almost no energy-dependence in a 6 MV beam. KCl:Eu²⁺ pellet samples showed a large dynamic range from 0.01 cGy to 60 Gy dose-to-water, and saturated at approximately 500 Gy as a result of KCl's intrinsic high radiation hardness. Taken together, this work provides strong evidence that KCl:Eu²⁺-based SPF with associated readout apparatus could result in a novel electronic film system that has all the desirable features associated with classic radiographic film and, importantly, water equivalence and the capability of permanent identification of each detector.

摘要

相似文献

1
Two-dimensional high spatial-resolution dosimeter using europium doped potassium chloride: a feasibility study.
Phys Med Biol. 2014 Apr 21;59(8):1899-909. doi: 10.1088/0031-9155/59/8/1899. Epub 2014 Mar 20.
4
Performance of KCl:Eu2+ storage phosphor dosimeters for low-dose measurements.
Phys Med Biol. 2013 Jun 21;58(12):4357-66. doi: 10.1088/0031-9155/58/12/4357. Epub 2013 Jun 5.
5
Quantitative proton radiation therapy dosimetry using the storage phosphor europium-doped potassium chloride.
Med Phys. 2020 Oct;47(10):5287-5300. doi: 10.1002/mp.14423. Epub 2020 Aug 16.
7
Temporal signal stability of KCl:Eu(2+) storage phosphor dosimeters.
Med Phys. 2013 Feb;40(2):021721. doi: 10.1118/1.4789482.
9
Dual-storage phosphor proton therapy dosimetry: Simultaneous quantification of dose and linear energy transfer.
Med Phys. 2021 Apr;48(4):1941-1955. doi: 10.1002/mp.14748. Epub 2021 Feb 19.

引用本文的文献

本文引用的文献

1
Storage Phosphors for Medical Imaging.
Materials (Basel). 2011 Jun 9;4(6):1034-1086. doi: 10.3390/ma4061034.
2
X ray storage performance of KCl:Eu with high cumulated dose.
Nucl Instrum Methods Phys Res B. 2014 May 1;326:178-181. doi: 10.1016/j.nimb.2013.08.067.
4
Performance of KCl:Eu2+ storage phosphor dosimeters for low-dose measurements.
Phys Med Biol. 2013 Jun 21;58(12):4357-66. doi: 10.1088/0031-9155/58/12/4357. Epub 2013 Jun 5.
6
Dosimetry tools and techniques for IMRT.
Med Phys. 2011 Mar;38(3):1313-38. doi: 10.1118/1.3514120.
7
Homogeneity of Gafchromic EBT2 film.
Med Phys. 2010 Apr;37(4):1753-6. doi: 10.1118/1.3368601.
10
TG-69: radiographic film for megavoltage beam dosimetry.
Med Phys. 2007 Jun;34(6):2228-58. doi: 10.1118/1.2736779.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验