Montei Carolyn, McDougal Susan, Mozola Mark, Rice Jennifer
J AOAC Int. 2014 Jan-Feb;97(1):155-8. doi: 10.5740/jaoacint.13-120.
The Soleris Non-fermenting Total Viable Count method was previously validated for a wide variety of food products, including cocoa powder. A matrix extension study was conducted to validate the method for use with cocoa butter and cocoa liquor. Test samples included naturally contaminated cocoa liquor and cocoa butter inoculated with natural microbial flora derived from cocoa liquor. A probability of detection statistical model was used to compare Soleris results at multiple test thresholds (dilutions) with aerobic plate counts determined using the AOAC Official Method 966.23 dilution plating method. Results of the two methods were not statistically different at any dilution level in any of the three trials conducted. The Soleris method offers the advantage of results within 24 h, compared to the 48 h required by standard dilution plating methods.
Soleris非发酵总活菌数方法先前已针对包括可可粉在内的多种食品进行了验证。开展了一项基质扩展研究,以验证该方法用于可可脂和可可液块的效果。测试样品包括天然污染的可可液块以及接种了源自可可液块的天然微生物菌群的可可脂。使用检测概率统计模型,将多个测试阈值(稀释度)下的Soleris结果与采用AOAC官方方法966.23稀释平板法测定的需氧平板计数进行比较。在进行的三项试验中的任何一项中,两种方法在任何稀释水平下的结果均无统计学差异。与标准稀释平板法所需的48小时相比,Soleris方法具有在24小时内得出结果的优势。