Suppr超能文献

High-performance liquid chromatographic separation of human haemoglobins. Simultaneous quantitation of foetal and glycated haemoglobins.

作者信息

Bisse E, Wieland H

机构信息

Central Laboratory, University Hospital, Frieburg, F.R.G.

出版信息

J Chromatogr. 1988 Dec 29;434(1):95-110. doi: 10.1016/0378-4347(88)80065-3.

Abstract

A high-performance liquid chromatographic system, which uses a weak cation exchanger (PolyCATA) together with Bis-Tris buffer (pH 6.47-7.0) and sodium acetate gradients, is described. Samples from adults and newborns were analysed and a clean separation of many minor and major normal and abnormal haemoglobin (Hb) variants was greatly improved. The method allows the separation of minor foetal haemoglobin (HbF) variants and the simultaneous quantitation of HbF and glycated HbA. HbF values correlated well with those obtained by the alkali denaturation method (r = 0.997). The glycated haemoglobin (HbAIc) levels measured in patients with high HbF concentrations correlated with the total glycated haemoglobin determined by bioaffinity chromatography (r = 0.973). The procedure is useful for diagnostic applications and affords an effective and sensitive way of examining blood samples for haemoglobin abnormalities.