Measuring level of agreement between values obtained by directly measured blood pressure and ultrasonic Doppler flow detector in cats.

作者信息

da Cunha Anderson F, Saile Katrin, Beaufrère Hugues, Wolfson Wendy, Seaton Diana, Acierno Mark J

机构信息

Department of Veterinary Clinical Science, School of Veterinary Medicine, Louisiana State University, Baton Rouge, LA, 70810.

出版信息

J Vet Emerg Crit Care (San Antonio). 2014 May-Jun;24(3):272-8. doi: 10.1111/vec.12161. Epub 2014 Apr 3.

Abstract

OBJECTIVE

To determine if blood pressure measured with an ultrasonic Doppler flow detector (Doppler) is in good agreement with directly measured blood pressures in anesthetized cats.

DESIGN

Prospective observational study.

SETTING

University veterinary teaching hospital.

ANIMALS

Thirty-nine cats undergoing routine neutering.

INTERVENTIONS

Cats were divided into 2 groups; 19 cats enrolled in Group A had a 24-Ga catheter inserted into a dorsal pedal artery; 20 cats in Group B had a 20-Ga catheter placed in a femoral artery. In both groups, systolic, diastolic, and mean arterial pressures were directly measured using a validated pressure measurement system. Indirect values were compared against direct blood pressure measurements.

RESULTS

There was no difference between groups. Overall, there was poor agreement with a significant bias observed between Doppler and directly measured blood pressures. For the systolic arterial pressure the bias was -8.8 with limits of agreements (LOA) of -39.3 and 21.7. For the mean arterial pressure, the bias was 14.0 with LOA of -13.9 and 41.9. For the diastolic arterial pressure, the bias was 27.9 with LOA of -4.4 and 60.2. Methodology, weight, sex, and replicates did not have a significant effect on the difference between indirect and direct measurements in any model.

CONCLUSIONS

Results suggest poor agreement between Doppler values and directly measured blood pressures in anesthetized cats. Use of Doppler in cats could be misleading and readings should be interpreted with caution in a clinical context.

摘要

目的

确定使用超声多普勒血流探测器(多普勒)测量的血压与麻醉猫直接测量的血压是否高度一致。

设计

前瞻性观察研究。

地点

大学兽医教学医院。

动物

39只接受常规绝育手术的猫。

干预措施

猫被分为两组;A组的19只猫将一根24G导管插入背侧足背动脉;B组的20只猫将一根20G导管置于股动脉。两组均使用经过验证的压力测量系统直接测量收缩压、舒张压和平均动脉压。将间接测量值与直接血压测量值进行比较。

结果

两组之间无差异。总体而言,多普勒测量值与直接测量的血压之间一致性较差,存在显著偏差。收缩压的偏差为-8.8,一致性界限(LOA)为-39.3和21.7。平均动脉压的偏差为14.0,LOA为-13.9和41.9。舒张压的偏差为27.9,LOA为-4.4和60.2。在任何模型中,方法、体重、性别和重复测量对间接测量值与直接测量值之间的差异均无显著影响。

结论

结果表明,麻醉猫的多普勒测量值与直接测量的血压之间一致性较差。在猫身上使用多普勒可能会产生误导,在临床环境中解读读数时应谨慎。