de la Vega J M, Ruiz-Arrebola S, Tornero-López A M, Vilches M, Guerrero R, Guirado D, Lallena A M
Servicio de Radiofísica, Hosp. Univ. "San Cecilio", Avda Dr. Olóriz, 16, E-18012 Granada, Spain.
Departamento de Oncología Radioterápica, Clínica Santa María, Avda Santa María, 0500, 7520378 Santiago de Chile, Chile.
Phys Med. 2014 Nov;30(7):827-32. doi: 10.1016/j.ejmp.2014.03.007. Epub 2014 Apr 13.
The relation between the data recorded with any device for the daily checking of the behavior of a clinical linac and the reference magnitudes to be monitored may be unknown. An experimental method relating the energy stability of the electron beam measured with StarTrack(®) to the R50 beam quality index is proposed. The bending magnet current is varied producing a change in the exit energy window and, therefore, a modification of the R50 value. For different values of this current, the output data of StarTrack(®) and the R50, obtained from depth doses measured in a water phantom are determined. A linear fit between both sets of data allows the identification of the StarTrack(®) output that provides the best way to obtain the quality index R50, for each beam nominal energy. Using these fits, an historical datum series is used to analyze the method proposed in the daily quality control. The ouput data of the StarTrack(®) and the R50 values show a good linear correlation. It is possible to establish a methodology that allows the monitoring of R50 by direct use of the daily quality control data measured with StarTrack(®). A method to monitor R50 in the daily quality control using the StarTrack(®) device has been developed. The method may be applied to similar devices in which the statistical control variable does not show a linear behavior with R50.
用于日常检查临床直线加速器运行情况的任何设备所记录的数据与待监测的参考量之间的关系可能是未知的。本文提出了一种实验方法,用于建立用StarTrack(®)测量的电子束能量稳定性与R50束流质量指标之间的关系。通过改变弯曲磁铁电流,使出射能量窗口发生变化,从而改变R50值。对于该电流的不同值,确定StarTrack(®)的输出数据以及从水模体中深度剂量测量获得的R50值。两组数据之间的线性拟合能够确定对于每个束流标称能量而言,哪种StarTrack(®)输出能以最佳方式获得质量指标R50。利用这些拟合结果,通过一个历史数据序列来分析日常质量控制中所提出的方法。StarTrack(®)的输出数据与R50值呈现出良好的线性相关性。有可能建立一种方法,通过直接使用StarTrack(®)测量的日常质量控制数据来监测R50。已经开发出一种在日常质量控制中使用StarTrack(®)设备监测R50的方法。该方法可应用于统计控制变量与R50不呈线性关系的类似设备。