Suppr超能文献

A CMOS ASIC Design for SiPM Arrays.

作者信息

Dey Samrat, Banks Lushon, Chen Shaw-Pin, Xu Wenbin, Lewellen Thomas K, Miyaoka Robert S, Rudell Jacques C

机构信息

Electrical Engineering Department, University of Washington, Seattle, WA 98195-2500, USA telephone: (206)685-1600.

Radiology Department, University of Washington, Seattle, WA 98105, USA telephone:(206)543-2084.

出版信息

IEEE Trans Nucl Sci. 2011 Dec;2011:732-737. doi: 10.1109/NSSMIC.2011.6154092.

Abstract

Our lab has previously reported on novel board-level readout electronics for an 8×8 silicon photomultiplier (SiPM) array featuring row/column summation technique to reduce the hardware requirements for signal processing. We are taking the next step by implementing a monolithic CMOS chip which is based on the row-column architecture. In addition, this paper explores the option of using diagonal summation as well as calibration to compensate for temperature and process variations. Further description of a timing pickoff signal which aligns all of the positioning (spatial channels) pulses in the array is described. The ASIC design is targeted to be scalable with the detector size and flexible to accommodate detectors from different vendors. This paper focuses on circuit implementation issues associated with the design of the ASIC to interface our Phase II MiCES FPGA board with a SiPM array. Moreover, a discussion is provided for strategies to eventually integrate all the analog and mixed-signal electronics with the SiPM, on either a single-silicon substrate or multi-chip module (MCM).

摘要

相似文献

1
A CMOS ASIC Design for SiPM Arrays.
IEEE Trans Nucl Sci. 2011 Dec;2011:732-737. doi: 10.1109/NSSMIC.2011.6154092.
2
Impact of Analog IC Impairments in SiPM Interface Electronics.
IEEE Trans Nucl Sci. 2012;2012:3572-3574. doi: 10.1109/NSSMIC.2012.6551818.
3
Highly-Integrated CMOS Interface Circuits for SiPM-Based PET Imaging Systems.
IEEE Nucl Sci Symp Conf Rec (1997). 2012:3556-3559. doi: 10.1109/nssmic.2012.6551814.
4
A CMOS Front-End Interface ASIC for SiPM-based Positron Emission Tomography Imaging Systems.
IEEE Biomed Circuits Syst Conf. 2017 Oct;2017. doi: 10.1109/BIOCAS.2017.8325059. Epub 2018 Mar 29.
5
Study of PET Detector Performance with Varying SiPM Parameters and Readout Schemes.
IEEE Trans Nucl Sci. 2011;58(3):590-596. doi: 10.1109/TNS.2011.2119378.
6
Evaluation of Matrix9 silicon photomultiplier array for small-animal PET.
Med Phys. 2015 Feb;42(2):585. doi: 10.1118/1.4905088.
7
Performance of long rectangular semi-monolithic scintillator PET detectors.
Med Phys. 2019 Apr;46(4):1608-1619. doi: 10.1002/mp.13432. Epub 2019 Feb 20.
8
Impact on the Spatial Resolution Performance of a Monolithic Crystal PET Detector Due to Different Sensor Parameters.
IEEE Nucl Sci Symp Conf Rec (1997). 2009 Oct 24;2009:3102-3105. doi: 10.1109/NSSMIC.2009.5401674.
9
Silicon photomultiplier signal readout and multiplexing techniques for positron emission tomography: a review.
Biomed Eng Lett. 2022 Jul 16;12(3):263-283. doi: 10.1007/s13534-022-00234-y. eCollection 2022 Aug.
10
An 8×8 Row-Column Summing Readout Electronics for Preclinical Positron Emission Tomography Scanners.
IEEE Nucl Sci Symp Conf Rec (1997). 2009 Oct 24;2009:2376-2380. doi: 10.1109/NSSMIC.2009.5402200.

引用本文的文献

1
A time-based single transmission-line readout with position multiplexing.
Biomed Eng Lett. 2022 Jan 17;12(1):85-95. doi: 10.1007/s13534-022-00215-1. eCollection 2022 Feb.
2
Impact of Analog IC Impairments in SiPM Interface Electronics.
IEEE Trans Nucl Sci. 2012;2012:3572-3574. doi: 10.1109/NSSMIC.2012.6551818.
3
Highly-Integrated CMOS Interface Circuits for SiPM-Based PET Imaging Systems.
IEEE Nucl Sci Symp Conf Rec (1997). 2012:3556-3559. doi: 10.1109/nssmic.2012.6551814.

本文引用的文献

1
Evolution of the Design of a Second Generation FireWire Based Data Acquisition System.
IEEE Nucl Sci Symp Conf Rec (1997). 2010 Oct 30:2510-2514. doi: 10.1109/NSSMIC.2010.5874239.
2
FPGA-Based Front-End Electronics for Positron Emission Tomography.
FPGA. 2009 Feb 22;2009(7):93-102. doi: 10.1145/1508128.1508143.
3
Design of an FPGA-Based Algorithm for Real-Time Solutions of Statistics-Based Positioning.
IEEE Trans Nucl Sci. 2010 Feb;57(1):71-77. doi: 10.1109/TNS.2009.2030581.
4
An 8×8 Row-Column Summing Readout Electronics for Preclinical Positron Emission Tomography Scanners.
IEEE Nucl Sci Symp Conf Rec (1997). 2009 Oct 24;2009:2376-2380. doi: 10.1109/NSSMIC.2009.5402200.
5
FPGA-Based Pulse Parameter Discovery for Positron Emission Tomography.
IEEE Nucl Sci Symp Conf Rec (1997). 2009 Oct 24;2009:2956-2961. doi: 10.1109/NSSMIC.2009.5401602.
6
Design of a Second Generation Firewire Based Data Acquisition System for Small Animal PET Scanners.
IEEE Nucl Sci Symp Conf Rec (1997). 2008 Oct 19;2008:5023-5028. doi: 10.1109/nssmic.2008.4774369.
7
Simulation of Algorithms for Pulse Timing in FPGAs.
IEEE Nucl Sci Symp Conf Rec (1997). 2007;4:3161-3165. doi: 10.1109/NSSMIC.2007.4436798.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验