Check-in and sorting of centrifuged serum and lithium-heparin tubes may be unsuitable using a bulk input module.

作者信息

Da Rin Giorgio, Lippi Giuseppe

机构信息

Department of Laboratory Medicine, Hospital of Bassano del Grappa, Bassano del Grappa (VI), Italy.

Laboratory of Clinical Chemistry and Hematology, Academic Hospital of Parma, Parma, Italy

出版信息

J Lab Autom. 2014 Oct;19(5):474-7. doi: 10.1177/2211068214536048. Epub 2014 May 27.

Abstract

The aim of this study was to establish that laboratory testing may be impaired by pouring and sorting of centrifuged tubes into a bulk input module. The values of 17 analytes, including albumin, aspartate aminotransferase (AST), bilirubin, urea nitrogen, C-reactive protein (CRP), calcium, chloride, cholesterol, creatine kinase (CK), creatinine, ferritin, glucose, lactate dehydrogenase (LDH), potassium, sodium, total protein, and thyroid-stimulating hormone (TSH), were measured in 29 centrifuged serum samples collected in tubes with gel separator and paired centrifuged lithium-heparin samples collected in tubes with gel separator, with (S-BIM) or without (S-NO-BIM) being poured and sorted by a bulk input module (Inpeco, Lugano, Switzerland). In serum tubes, significant differences were observed in S-BIM values of albumin, AST, cholesterol, ferritin, glucose, LDH, potassium, and sodium compared with S-NO-BIM, with bias of LDH exceeding the total allowable error. In lithium-heparin tubes, statistically significant differences were observed in S-BIM values of all parameters, except urea nitrogen, CRP, and sodium, compared with S-NO-BIM. The percentage bias of AST, LDH, glucose, and creatinine exceeded the total allowable error. These results demonstrate that check-in and sorting of centrifuged serum and lithium-heparin tubes may be unsuitable using a bulk input module.

摘要

本研究的目的是确定将离心管倒入并分拣到批量输入模块中可能会影响实验室检测。在29份收集于带凝胶分离器试管中的离心血清样本以及配对的收集于带凝胶分离器试管中的离心锂肝素样本中,测量了17种分析物的值,这些样本有(S-BIM)或没有(S-NO-BIM)通过批量输入模块(Inpeco,瑞士卢加诺)进行倒入和分拣。在血清试管中,与S-NO-BIM相比,白蛋白、天冬氨酸转氨酶(AST)、胆固醇、铁蛋白、葡萄糖、乳酸脱氢酶(LDH)、钾和钠的S-BIM值存在显著差异,LDH的偏差超过了总允许误差。在锂肝素试管中,与S-NO-BIM相比,除尿素氮、C反应蛋白(CRP)和钠外,所有参数的S-BIM值均存在统计学显著差异。AST、LDH、葡萄糖和肌酐的偏差百分比超过了总允许误差。这些结果表明,使用批量输入模块对离心血清和锂肝素试管进行登记和分拣可能不合适。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验