Niemann Tilo, Henry Simon, Duhamel Alain, Faivre Jean-Baptiste, Deschildre Antoine, Colas Lucie, Santangelo Teresa, Remy Jacques, Remy-Jardin Martine
Department of Thoracic Imaging, Hospital Calmette, Univ Lille Nord de France, Lille, France.
Pediatr Radiol. 2014 Nov;44(11):1347-57. doi: 10.1007/s00247-014-3027-8. Epub 2014 Jun 13.
Before introducing 70-kVp settings in the low-kilovoltage strategies for pediatric examinations, it was mandatory to demonstrate, at similar dose levels, an equivalence of image quality at 70 kVp and 80 kVp.
To assess image quality of chest CT examinations acquired at 70 kVp in comparison with standard scanning at 80 kVp.
We prospectively evaluated 129 children with a 70-kVp scanning protocol (group 1). All scanning parameters were kept similar to those usually selected for pediatric standard 80-kVp protocols, except the milliamperage increased by a factor of 1.6 to maintain comparable radiation dose. Image quality of group 1 examinations was compared to that of a paired population scanned at 80 kVp (group 2). The noninferiority hypothesis was fixed at 10% of the mean level of image noise.
There was no significant difference in the mean dose length product (DLP) and the volume computed tomography dose index (CTDIvol) between the groups (DLP: 20.5 ± 5.8 mGy.cm [group 1] vs. 19.7 ± 7.6 mGy.cm [group 2]; P = 0.06) (CTDIvol: 0.8 ± 0.1 mGy [group 1] vs. 0.8 ± 0.18 mGy [group 2]; P = 0.94). The mean of differences in image noise between group 1 and group 2 examinations was -1.38 (-2.59; -0.18), verifying the noninferiority hypothesis. Subjective image quality did not significantly differ between group 1 and group 2 examinations (P = 0.18).
At equivalent radiation dose levels, 70-kVp protocols provide similar image quality to that achievable at 80 kVp.
在将70 kVp设置引入儿科检查的低千伏策略之前,必须在相似剂量水平下证明70 kVp和80 kVp时图像质量的等效性。
评估70 kVp下胸部CT检查的图像质量,并与80 kVp的标准扫描进行比较。
我们前瞻性地评估了129名采用70 kVp扫描方案的儿童(第1组)。除毫安数增加1.6倍以维持相当的辐射剂量外,所有扫描参数均保持与儿科标准80 kVp方案通常选择的参数相似。将第1组检查的图像质量与在80 kVp下扫描的配对人群(第2组)的图像质量进行比较。非劣效性假设设定为图像噪声平均水平的10%。
两组之间的平均剂量长度乘积(DLP)和容积计算机断层扫描剂量指数(CTDIvol)无显著差异(DLP:第1组为20.5±5.8 mGy·cm,第2组为19.7±7.6 mGy·cm;P = 0.06)(CTDIvol:第1组为0.8±0.1 mGy,第2组为0.8±0.18 mGy;P = 0.94)。第1组和第2组检查之间图像噪声差异的平均值为-1.38(-2.59;-0.18),验证了非劣效性假设。第1组和第2组检查之间的主观图像质量无显著差异(P = 0.18)。
在等效辐射剂量水平下,70 kVp方案提供的图像质量与80 kVp时可达到的图像质量相似。