Performance evaluation of a dual-energy x-ray bone densitometer.

作者信息

Mazess R, Collick B, Trempe J, Barden H, Hanson J

机构信息

LUNAR Radiation Corporation, Madison, Wisconsin 53713.

出版信息

Calcif Tissue Int. 1989 Mar;44(3):228-32. doi: 10.1007/BF02556569.

Abstract

We tested a dual-energy bone densitometer (LUNAR DPX) that uses a stable x-ray generator and a K-edge filter to achieve the two energy levels. A conventional scintillation detector in pulse-counting mode was used together with a gain stabilizer. The densitometer normally performs spine and femur scans in about 6 minutes and 3 minutes, respectively, with adequate spatial resolution (1.2 x 1.2 mm). Total body scans take either 10 minutes or 20 minutes. The long-term (6 months, n = 195) precision of repeat measurement on an 18-cm thick spine phantom was 0.6% at the medium speed. Precision error in vivo was about 0.6, 0.9 and 1.5% for spine scans (L2-L4) at slow, medium and fast speeds, while the error was 1.2 and 1.5 to 2.0%, respectively, for femur scans at slow and medium speed. The precision of total body bone density was 0.5% in vitro and in vivo. The response to increasing amounts of calcium hydroxyapatite was linear (r = 0.99). The densitometer accurately indicated (within 1%) the actual amount of hydroxyapatite after correction for physiological amounts of marrow fat. The measured area corresponded exactly (within 0.5%) to that of known annuli and to the radiographic area of spine phantoms. There was no significant effect of tissue thickness on mass, area, or areal density (BMD) between 10 and 24 cm of water. The BMD values for both spine and femur in vivo correlated highly (r = 0.98, SEE = .03 g/cm2) with those obtained using conventional 153Gd DPA. Similarly, total body BMD correlated highly (r = 0.96, SEE = .02 g/cm2) with DPA results.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验