Parati G, Casadei R, Groppelli A, Di Rienzo M, Mancia G
Cattedra di Semeiotica Medica, Università di Milano, Italy.
Hypertension. 1989 Jun;13(6 Pt 1):647-55. doi: 10.1161/01.hyp.13.6.647.
The accuracy of blood pressure values obtained by continuous noninvasive finger blood pressure recording via the FINAPRES device was evaluated by comparison with simultaneous intraarterial monitoring both at rest and during performance of tests known to induce fast and often marked changes in blood pressure. The comparison was performed in 24 normotensive or essential hypertensive subjects. The average discrepancy between finger and intra-arterial blood pressure recorded over a 30-minute rest period was 6.5 +/- 2.6 mm Hg and 5.4 +/- 2.9 mm Hg for systolic and diastolic blood pressure, respectively; a close between-method correspondence was also demonstrated by linear regression analysis. The beat-to-beat changes in finger systolic and diastolic blood pressure were on average similar to those measured intra-arterially during tests that induced a pressor or depressor response (hand-grip, cold pressor test, diving test, Valsalva maneuver, intravenous injections of phenylephrine and trinitroglycerine) as well as during tests that caused vasomotor changes without major variations in blood pressure (application of lower body negative pressure, passive leg raising). The average between-method discrepancy in the evaluation of blood pressure changes was never greater than 4.3 and 2.0 mm Hg for systolic and diastolic blood pressure, respectively; the corresponding standard deviations ranged between 4.6 and 1.6 mm Hg. Beat-to-beat computer analysis of blood pressure variability over the 30-minute rest period provided standard deviations almost identical when calculated by separate consideration of intra-arterial and finger blood pressure tracings (3.7 and 3.8 mm Hg, respectively).(ABSTRACT TRUNCATED AT 250 WORDS)
通过比较静息状态以及进行已知会引起血压快速且通常显著变化的测试期间,使用FINAPRES设备连续无创记录手指血压所获得的血压值的准确性,与同时进行的动脉内监测结果进行了评估。该比较在24名血压正常或原发性高血压受试者中进行。在30分钟的静息期内,记录的手指血压与动脉内血压之间的平均差异,收缩压为6.5±2.6毫米汞柱,舒张压为5.4±2.9毫米汞柱;线性回归分析也表明了两种方法之间具有密切的对应关系。在诱导升压或降压反应的测试(握力、冷加压试验、潜水试验、瓦尔萨尔瓦动作、静脉注射去氧肾上腺素和三硝酸甘油)期间,以及在引起血管舒缩变化但血压无重大变化的测试(下体负压应用、被动抬腿)期间,手指收缩压和舒张压的逐搏变化平均与动脉内测量的变化相似。在评估血压变化时,两种方法之间的平均差异,收缩压和舒张压分别从不超过4.3和2.0毫米汞柱;相应的标准差在4.6和1.6毫米汞柱之间。在30分钟的静息期内,通过分别考虑动脉内和手指血压描记图计算的血压变异性的逐搏计算机分析提供了几乎相同的标准差(分别为3.7和3.8毫米汞柱)。(摘要截选至250字)