Suppr超能文献

连续无创血压监测:Valsalva动作期间Finapres设备的可靠性

Continuous non-invasive blood pressure monitoring: reliability of Finapres device during the Valsalva manoeuvre.

作者信息

Imholz B P, van Montfrans G A, Settels J J, van der Hoeven G M, Karemaker J M, Wieling W

机构信息

Department of Internal Medicine, University of Amsterdam, The Netherlands.

出版信息

Cardiovasc Res. 1988 Jun;22(6):390-7. doi: 10.1093/cvr/22.6.390.

Abstract

Because of the inherent risks of intra-arterial blood pressure monitoring a new non-invasive device, Finapres, which measures blood pressure continuously in the finger, was evaluated in 14 hypertensive and one normotensive subject. Brachial intra-arterial and finger pressures were compared during a control period and a subsequent Valsalva manoeuvre. Visually, blood pressures measured by Finapres faithfully reproduced the intra-arterial recordings in all subjects. From each pressure signal beat to beat systolic, diastolic, and mean blood pressure values and their differences were obtained and the time course of the response and its characteristic features were analysed. During the control period the Finapres measurements were lower than intra-arterial systolic, mean, and diastolic pressures (mean(SD) 1(9.6), 9(6.8), and 4(6.1) mmHg respectively). During the response to the Valsalva manoeuvre the brachial-finger pressure differences showed limited deviation from those during the control period; median differences were at most 6 mmHg occurring late during the intrathoracic strain period and 7 mmHg during the post-release blood pressure overshoot. In general, the Finapres device reproduced intra-arterial patterns faithfully. This device appears to offer a reliable alternative to intra-arterial blood pressure monitoring.

摘要

由于动脉内血压监测存在固有风险,因此对一种新的非侵入性设备——Finapres进行了评估,该设备可连续测量手指血压。研究纳入了14名高血压患者和1名血压正常的受试者。在对照期和随后的瓦尔萨尔瓦动作期间,对比了肱动脉内血压和手指血压。从视觉上看,Finapres测量的血压在所有受试者中都如实地再现了动脉内记录的血压。从每个压力信号中获取逐搏收缩压、舒张压和平均血压值及其差值,并分析反应的时间过程及其特征。在对照期,Finapres测量值低于动脉内收缩压、平均压和舒张压(分别为平均(标准差)1(9.6)、9(6.8)和4(6.1)mmHg)。在对瓦尔萨尔瓦动作的反应过程中,肱动脉与手指的压力差值与对照期相比偏差有限;在胸腔内压力增加期后期,中位数差值最大为6 mmHg,在血压释放后过冲期为7 mmHg。总体而言,Finapres设备如实地再现了动脉内血压模式。该设备似乎为动脉内血压监测提供了一种可靠的替代方法。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验