The development and validation of an early warning system to prevent the acquisition of 12-lead resting ECGs with interchanged electrode positions.

作者信息

de Bie Johan, Mortara David W, Clark Todd F

机构信息

Mortara Instrument Inc., 7865 North 86th Street, 53224 Milwaukee, WI, USA.

Mortara Instrument Inc., 7865 North 86th Street, 53224 Milwaukee, WI, USA.

出版信息

J Electrocardiol. 2014 Nov-Dec;47(6):794-7. doi: 10.1016/j.jelectrocard.2014.08.015. Epub 2014 Aug 23.

Abstract

BACKGROUND

A fraction of routine resting ECG's are taken with electrode positions interchanged, leading to possible clinical misinterpretation.

OBJECTIVE

Develop and test a method to detect and prevent electrode reversals at the electrocardiograph before the ECG is acquired.

METHOD

The algorithm is based on QRS axis and P amplitudes for limb electrode reversals, and P-Q-RS amplitude distances to detect chest electrode reversals. The evaluation method involved a large (>18,000) hospital database for which serial ECG's were available and was based on simulated juxtapositions.

RESULTS

The 7 most common lead reversals could be detected with a specificity of 99.8% per type and an average sensitivity of 90%, excluding LA-LL reversal (22% sensitivity).

DISCUSSION

Results are similar to retrospective studies that used smaller, more homogeneous datasets.

CONCLUSION

The early warning system reduces the ECG's recorded with reversal by 80%, at the price of a modest false alert rate of 1.4%.