Chen David, Sharif Behzad, Dharmakumar Rohan, Thomson Louise E J, Bairey Merz C Noel, Berman Daniel S, Li Debiao
Department of Biomedical Engineering, Northwestern University, Chicago, Illinois, USA.
Biomedical Imaging Research Institute, Department of Biomedical Sciences, Cedars-Sinai Medical Center, Los Angeles, California, USA.
Magn Reson Med. 2015 Sep;74(3):765-71. doi: 10.1002/mrm.25451. Epub 2014 Sep 16.
MR myocardial perfusion imaging is dependent on reliable electrocardiogram (ECG) triggering for accurate measurement of myocardial blood flow (MBF). A non-ECG-triggered method for quantitative first-pass imaging may improve clinical feasibility in patients with poor ECG signal. The purpose of this study is to evaluate the feasibility of a non-ECG-triggered method for myocardial perfusion imaging in a single slice.
The proposed non-ECG-triggered technique uses a saturation-recovery magnetization preparation and golden-angle radial acquisition for integrated arterial input function (AIF) measurement. Image based self-gating with a temporal resolution of 42.6 ms is used to generate a first-pass image series with consistent cardiac phase. The AIF is measured using beat-by-beat T1 estimation of the ventricular blood pool. The proposed technique was performed on 14 healthy volunteers and compared against a conventional ECG-triggered dual-bolus acquisition.
The proposed method produced MBF with no significant difference compared with ECG-triggered technique (mean of 0.63 ± 0.22 mL/min/g to 0.73 ± 0.21 mL/min/g).
We have developed a non-ECG-triggered perfusion imaging method with T1 based measurement of the AIF in a single slice. In this preliminary study, our results demonstrate that MBF measured using the proposed method is comparable to the conventional ECG-triggered method.
磁共振心肌灌注成像依赖可靠的心电图(ECG)触发来准确测量心肌血流量(MBF)。一种用于定量首过成像的非ECG触发方法可能会提高心电图信号不佳患者的临床可行性。本研究的目的是评估在单一层面进行心肌灌注成像的非ECG触发方法的可行性。
所提出的非ECG触发技术使用饱和恢复磁化准备和黄金角径向采集来测量综合动脉输入函数(AIF)。基于图像的自门控技术,时间分辨率为42.6毫秒,用于生成具有一致心动周期相位的首过图像序列。通过逐搏估计心室血池的T1来测量AIF。该技术在14名健康志愿者身上进行,并与传统的ECG触发双团注采集方法进行比较。
与ECG触发技术相比,所提出的方法得出的MBF无显著差异(平均值从0.63±0.22毫升/分钟/克到0.73±0.21毫升/分钟/克)。
我们开发了一种非ECG触发的灌注成像方法,在单一层面基于T1测量AIF。在这项初步研究中,我们的结果表明,使用所提出的方法测量的MBF与传统的ECG触发方法相当。