双能 X 射线吸收法测定中的错误发生率和类型。

Prevalence and type of errors in dual-energy x-ray absorptiometry.

机构信息

Scuola di Specializzazione in Radiodiagnostica, Università degli Studi di Milano, via Festa del Perdono 7, 20122, Milano, Italy,

出版信息

Eur Radiol. 2015 May;25(5):1504-11. doi: 10.1007/s00330-014-3509-y. Epub 2014 Nov 27.

Abstract

OBJECTIVES

Pitfalls in dual-energy x-ray absorptiometry (DXA) are common. Our aim was to assess rate and type of errors in DXA examinations/reports, evaluating a consecutive series of DXA images of patients examined elsewhere and later presenting to our institution for a follow-up DXA.

METHODS

After ethics committee approval, a radiologist retrospectively reviewed all DXA images provided by patients presenting at our institution for a new DXA. Errors were categorized as patient positioning (PP), data analysis (DA), artefacts and/or demographics.

RESULTS

Of 2,476 patients, 1,198 had no previous DXA, while 793 had a previous DXA performed in our institution. The remaining 485 (20 %) patients entered the study (38 men and 447 women; mean age ± standard deviation, 68 ± 9 years). Previous DXA examinations were performed at a total of 37 centres. Of 485 reports, 451 (93 %) had at least one error out of a total of 558 errors distributed as follows: 441 (79 %) were DA, 66 (12 %) PP, 39 (7 %) artefacts and 12 (2 %) demographics.

CONCLUSIONS

About 20 % of patients did not undergo DXA at the same institution as previously. More than 90 % of DXA presented at least one error, mainly of DA. International Society for Clinical Densitometry guidelines are very poorly adopted.

KEY POINTS

• More than 90 % of DXA examinations/reports presented one or more errors. • About 80 % of errors are related to image data analysis. • Errors in DXA examinations may have potential implications for patients' management.

摘要

目的

双能 X 射线吸收法(DXA)中的陷阱很常见。我们的目的是评估在其他地方检查后到我院进行后续 DXA 的患者的 DXA 检查/报告中的错误发生率和类型。

方法

在伦理委员会批准后,一位放射科医生回顾性地审查了我院新进行的 DXA 检查的所有患者的 DXA 图像。错误被分为患者定位(PP)、数据分析(DA)、伪影和/或人口统计学。

结果

在 2476 名患者中,1198 名患者之前没有进行过 DXA,而 793 名患者之前在我院进行过 DXA。其余 485 名(20%)患者进入研究(38 名男性和 447 名女性;平均年龄±标准差,68±9 岁)。之前的 DXA 检查是在总共 37 个中心进行的。在 485 份报告中,有 451 份(93%)有 558 个错误中的至少一个,分布如下:441 个(79%)为 DA,66 个(12%)为 PP,39 个(7%)为伪影,12 个(2%)为人口统计学。

结论

大约 20%的患者之前没有在同一机构进行 DXA。超过 90%的 DXA 检查报告至少有一个错误,主要是 DA。国际临床密度测量学协会指南的采用情况非常不理想。

关键点

•超过 90%的 DXA 检查/报告存在一个或多个错误。•约 80%的错误与图像数据分析有关。•DXA 检查中的错误可能对患者的管理产生潜在影响。