Ficko Bradley W, Nadar Priyanka M, Diamond Solomon G
Thayer School of Engineering at Dartmouth, 14 Engineering Dr. Hanover, New Hampshire, USA, 03755.
J Magn Magn Mater. 2015 Feb 1;375:164-176. doi: 10.1016/j.jmmm.2014.10.011.
This study demonstrates a method for alternating current (AC) susceptibility imaging (ASI) of magnetic nanoparticles (mNPs) using low cost instrumentation. The ASI method uses AC magnetic susceptibility measurement to create tomographic images using an array of drive coils, compensation coils and fluxgate magnetometers. Using a spectroscopic approach in conjunction with ASI, a series of tomographic images can be created for each frequency measurement and is termed sASI. The advantage of sASI is that mNPs can be simultaneously characterized and imaged in a biological medium. System calibration was performed by fitting the in-phase and out-of-phase susceptibility measurements of an mNP sample with a hydrodynamic diameter of 100 nm to a Brownian relaxation model (R = 0.96). Samples of mNPs with core diameters of 10 and 40 nm and a sample of 100 nm hydrodynamic diameter were prepared in 0.5 ml tubes. Three mNP samples were arranged in a randomized array and then scanned using sASI with six frequencies between 425 and 925 Hz. The sASI scans showed the location and quantity of the mNP samples (R = 0.97). Biological compatibility of the sASI method was demonstrated by scanning mNPs that were injected into a pork sausage. The mNP response in the biological medium was found to correlate with a calibration sample (R = 0.97, p <0.001). These results demonstrate the concept of ASI and advantages of sASI.
本研究展示了一种使用低成本仪器对磁性纳米颗粒(mNPs)进行交流(AC)磁化率成像(ASI)的方法。ASI方法利用交流磁化率测量,通过驱动线圈、补偿线圈和磁通门磁力计阵列创建断层图像。结合光谱方法与ASI,可为每个频率测量创建一系列断层图像,这被称为sASI。sASI的优势在于可以在生物介质中同时对mNPs进行表征和成像。通过将流体动力学直径为100 nm的mNP样品的同相和异相磁化率测量值拟合到布朗弛豫模型(R = 0.96)来进行系统校准。将核心直径为10和40 nm的mNP样品以及一个流体动力学直径为100 nm的样品制备在0.5 ml的试管中。将三个mNP样品以随机阵列排列,然后使用sASI在425至925 Hz之间的六个频率下进行扫描。sASI扫描显示了mNP样品的位置和数量(R = 0.97)。通过对注入猪肉香肠中的mNPs进行扫描,证明了sASI方法的生物相容性。发现生物介质中的mNP响应与校准样品相关(R = 0.97,p <0.001)。这些结果证明了ASI的概念以及sASI的优势。