Becher Tobias H, Bui Simon, Zick Günther, Bläser Daniel, Schädler Dirk, Weiler Norbert, Frerichs Inéz
Department of Anesthesiology and Intensive Care Medicine, University Medical Center Schleswig-Holstein, Campus Kiel, Arnold-Heller-Strasse 3, Haus 12, 24105, Kiel, Germany.
Crit Care. 2014 Dec 10;18(6):679. doi: 10.1186/s13054-014-0679-6.
Assessment of respiratory system compliance (Crs) can be used for individual optimization of positive end-expiratory pressure (PEEP). However, in patients with spontaneous breathing activity, the conventional methods for Crs measurement are inaccurate because of the variable muscular pressure of the patient. We hypothesized that a PEEP wave maneuver, analyzed with electrical impedance tomography (EIT), might be suitable for global and regional assessment of Crs during assisted spontaneous breathing.
After approval of the local ethics committee, we performed a pilot clinical study in 18 mechanically ventilated patients (61 ± 16 years (mean ± standard deviation)) who were suitable for weaning with pressure support ventilation (PSV). For the PEEP wave, PEEP was elevated by 1 cmH2O after every fifth breath during PSV. This was repeated five times, until a total PEEP increase of 5 cmH2O was reached. Subsequently, PEEP was reduced in steps of 1 cmH2O in the same manner until the original PEEP level was reached. Crs was calculated using EIT from the global, ventral and dorsal lung regions of interest. For reference measurements, all patients were also examined during controlled mechanical ventilation (CMV) with a low-flow pressure-volume maneuver. Global and regional Crs(low-flow) was calculated as the slope of the pressure-volume loop between the pressure that corresponded to the selected PEEP and PEEP +5 cmH2O. For additional reference, Crs during CMV (Crs(CMV)) was calculated as expired tidal volume divided by the difference between airway plateau pressure and PEEP.
Respiratory system compliance calculated from the PEEP wave (Crs(PEEP wave)) correlated closely with both reference measurements (r = 0.79 for Crs(low-flow) and r = 0.71 for Crs(CMV)). No significant difference was observed between the mean Crs(PEEP wave) and the mean Crs(low-flow). However, a significant bias of +17.1 ml/cmH2O was observed between Crs(PEEP wave) and Crs(CMV).
Analyzing a PEEP wave maneuver with EIT allows calculation of global and regional Crs during assisted spontaneous breathing. In mechanically ventilated patients with spontaneous breathing activity, this method might be used for assessment of the global and regional mechanical properties of the respiratory system.
评估呼吸系统顺应性(Crs)可用于呼气末正压(PEEP)的个体化优化。然而,在有自主呼吸活动的患者中,由于患者肌肉压力的变化,传统的Crs测量方法并不准确。我们推测,用电阻抗断层扫描(EIT)分析PEEP波操作可能适用于辅助自主呼吸期间Crs的整体和区域评估。
经当地伦理委员会批准后,我们对18例适合压力支持通气(PSV)撤机的机械通气患者(61±16岁(平均值±标准差))进行了一项初步临床研究。对于PEEP波,在PSV期间每五次呼吸后将PEEP升高1 cmH2O。重复此操作五次,直至PEEP总共增加5 cmH2O。随后,以相同方式将PEEP以1 cmH2O的步长降低,直至达到原始PEEP水平。使用EIT从感兴趣的全肺、肺腹侧和背侧区域计算Crs。为了进行参考测量,所有患者还在控制机械通气(CMV)期间通过低流量压力-容积操作进行了检查。将全肺和区域Crs(低流量)计算为对应于选定PEEP和PEEP +5 cmH2O之间压力-容积环的斜率。作为额外参考,将CMV期间的Crs(Crs(CMV))计算为呼出潮气量除以气道平台压与PEEP之间的差值。
根据PEEP波计算的呼吸系统顺应性(Crs(PEEP波))与两种参考测量值密切相关(Crs(低流量)的r = 0.79,Crs(CMV)的r = 0.71)。Crs(PEEP波)的平均值与Crs(低流量)的平均值之间未观察到显著差异。然而,Crs(PEEP波)与Crs(CMV)之间观察到+17.1 ml/cmH2O的显著偏差。
用EIT分析PEEP波操作可在辅助自主呼吸期间计算全肺和区域Crs。在有自主呼吸活动的机械通气患者中,该方法可用于评估呼吸系统的整体和区域力学特性。