Babaei A, Lin E C, Szabo A, Massey B T
Division of Gastroenterology and Hepatology, Medical College of Wisconsin, Milwaukee, WI, USA.
Neurogastroenterol Motil. 2015 Feb;27(2):277-84. doi: 10.1111/nmo.12493. Epub 2014 Dec 29.
Manoscan(™) is one of the commonly used high-resolution manometry (HRM) systems with declared measurement accuracy of 1-2 mmHg. However, the accuracy of pressure measurements is limited by development of pressure drift (PD) throughout recording. To date, there has been no systematic investigation to identify the factors contributing to PD. The aim of the present study was to characterize the frequency and magnitude of PD in Manoscan(™) system and identify the factors contributing to PD.
Records of 560 consecutive clinical esophageal HRM studies recorded by six distinct HRM catheters were retrospectively reviewed. PD was defined as the residual pressure measurement by each sensor immediately after removal of the catheter. Non-parametric locally weighted regression analysis was performed to assess the effect of duration of study, number of prior uses of a catheter, peak and average pressure exposure during a study on the PD.
The majority (95%) of clinical manometry studies showed a non-negligible PD of more than 5 mmHg. The overall PD was 13 ± 5 mmHg and the sensor with greatest amount of PD showed 23 ± 12 mmHg of drift. The upper esophageal sphincter showed the highest PD. Average pressure exposure of a sensor throughout the recording was the most important predictor of PD. PD inversely correlated with number of prior uses of a catheter.
CONCLUSIONS & INFERENCES: The PD preferentially affects esophageal high-pressure zones, and strongly correlates with 'average pressure exposure' of a sensor during manometry. Available algorithms of the analysis software do not adequately correct the PD.
Manoscan(™)是常用的高分辨率测压(HRM)系统之一,宣称测量精度为1 - 2 mmHg。然而,压力测量的准确性受到整个记录过程中压力漂移(PD)的影响。迄今为止,尚未有系统研究来确定导致PD的因素。本研究的目的是描述Manoscan(™)系统中PD的频率和幅度,并确定导致PD的因素。
回顾性分析了由六个不同的HRM导管记录的560例连续临床食管HRM研究记录。PD定义为导管移除后每个传感器的残余压力测量值。进行非参数局部加权回归分析,以评估研究持续时间、导管先前使用次数、研究期间的峰值和平均压力暴露对PD的影响。
大多数(95%)临床测压研究显示出超过5 mmHg的不可忽略的PD。总体PD为13±5 mmHg,PD量最大的传感器显示出23±12 mmHg的漂移。食管上括约肌显示出最高的PD。整个记录过程中传感器的平均压力暴露是PD的最重要预测因素。PD与导管先前使用次数呈负相关。
PD优先影响食管高压区,并且与测压期间传感器的“平均压力暴露”密切相关。分析软件的现有算法不能充分校正PD。