Suppr超能文献

Rapid culture confirmation of adenovirus isolates using the adenoclone enzyme immunoassay (EIA) test kit.

作者信息

Lipson S M

机构信息

Department of Pathology and Laboratories, Nassau County Medical Center, East Meadow, New York 11554.

出版信息

Diagn Microbiol Infect Dis. 1989 Jul-Aug;12(4):361-2. doi: 10.1016/0732-8893(89)90104-1.

Abstract

The adenoclone enzyme immunoassay (EIA) test kit (Cambridge BioScience Corp., Worcester, MA) was evaluated using 43 adenovirus and 53 nonadenovirus (i.e., polio-, echo-, coxsackievirus types A,B, parainfluenza virus, and herpes simplex virus) isolates obtained during routine clinical diagnostic screening procedures. Forty-two of 43 confirmed adenovirus isolates were reactive upon testing by the adenoclone test kit. None of the 53 nonadenovirus isolates, including seven difficult-to-cultivate group A coxsackievirus isolates, cross-reacted with the EIA kit. The adenoclone EIA kit is an appropriate supplement to the regimen of clinical diagnostic procedures for the culture confirmation of adenovirus.