Physikalisch-Chemisches Institut, Universität Zürich, Winterthurerstrasse 190, CH-8057 Zürich, Switzerland; E-Mails:
Biosensors (Basel). 2013 Feb 8;3(1):108-15. doi: 10.3390/bios3010108. eCollection 2013 Mar.
We present an immunoassay for Interferon-γ (IFN-γ) with a limit of detection of 1.9 pM (30 pg/mL) and a linear concentration range spanning three orders of magnitude. The developed one-step assay takes only 12 min and can replace the time-consuming and labor-intensive enzyme-linked immunosorbent assay (ELISA). The solid-phase sandwich assay is performed on a new measurement system comprising single-use test tubes and a compact fluorescence reader. The polymer tubes contain an optical configuration for the detection of supercritical angle fluorescence, allowing for highly sensitive real-time binding measurements.