可重复使用的 JuniorSTAR® 半单位胰岛素笔的功能评估

Functional Evaluation of the Reusable JuniorSTAR® Half-Unit Insulin Pen.

作者信息

Klonoff David, Nayberg Irina, Rabbone Ivana, Domenger Catherine, Stauder Udo, Oualali Hamid, Danne Thomas

机构信息

Diabetes Research Institute, Mills-Peninsula Health Services, San Mateo, CA, USA

Diabetes Research Institute, Mills-Peninsula Health Services, San Mateo, CA, USA.

出版信息

J Diabetes Sci Technol. 2015 May;9(3):625-31. doi: 10.1177/1932296815569246. Epub 2015 Jan 28.

Abstract

BACKGROUND

The functional performance of the JuniorSTAR(®) (Sanofi, Paris, France) half-unit insulin pen was evaluated through a series of specific objective tests to assess the dose accuracy, pen weight, injection force, and dialing torque.

METHOD

Pens (n = 60) were tested under standard atmospheric conditions with 3 different types of insulins manufactured by Sanofi (insulin glargine, insulin glulisine, and biphasic insulin isophane). The dose accuracy was tested according to the ISO 11608-1:2012 standards. Injection doses of 0.010, 0.155, and 0.300 ml were evaluated. For mean weight evaluation, the pens without the cartridge were weighed on precision balances. The injection force was measured using a texture analyzer and the dialing torque was measured using a torque meter.

RESULTS

JuniorSTAR met the ISO 11608-1:2012 criteria for dose accuracy as all the delivered doses were within the predefined limits for all types of insulin tested. The mean weight of the JuniorSTAR pen was 33.4 g (SD = 0.075). The mean injection force was 6.0 N (SD = 0.8), 4.3 N (SD = 0.4), and 5.1 N (SD = 0.6) for insulin glargine, insulin glulisine, and biphasic insulin isophane, respectively. The mean dialing torque was 5.09 Ncm (SD = 0.29) and 5.88 Ncm (SD = 0.53) for setting and correcting a dose, respectively.

CONCLUSIONS

Together with results from a previously reported usability survey, these results show that the JuniorSTAR reusable, half-unit pen is a lightweight and accurate device for insulin delivery with a dialing torque and injection force suitable for young people with type 1 diabetes.

摘要

背景

通过一系列特定的客观测试,对JuniorSTAR(®)(赛诺菲,法国巴黎)半单位胰岛素笔的功能性能进行了评估,以评估剂量准确性、笔的重量、注射力和调节扭矩。

方法

在标准大气条件下,对60支笔进行测试,使用赛诺菲生产的3种不同类型的胰岛素(甘精胰岛素、门冬胰岛素和低精蛋白锌胰岛素)。根据ISO 11608-1:2012标准测试剂量准确性。评估了0.010、0.155和0.300毫升的注射剂量。对于平均重量评估,将没有笔芯的笔放在精密天平上称重。使用质地分析仪测量注射力,使用扭矩计测量调节扭矩。

结果

JuniorSTAR符合ISO 11608-1:2012剂量准确性标准,因为所有测试的胰岛素类型的所有输送剂量均在预定义范围内。JuniorSTAR笔的平均重量为33.4克(标准差=0.075)。甘精胰岛素、门冬胰岛素和低精蛋白锌胰岛素的平均注射力分别为6.0牛(标准差=0.8)、4.3牛(标准差=0.4)和5.1牛(标准差=0.6)。设置和校正剂量时的平均调节扭矩分别为5.09牛厘米(标准差=0.29)和5.88牛厘米(标准差=0.53)。

结论

结合先前报告的可用性调查结果,这些结果表明,JuniorSTAR可重复使用的半单位笔是一种轻便且准确的胰岛素输送装置,其调节扭矩和注射力适合1型糖尿病青少年。

相似文献

1
Functional Evaluation of the Reusable JuniorSTAR® Half-Unit Insulin Pen.可重复使用的 JuniorSTAR® 半单位胰岛素笔的功能评估
J Diabetes Sci Technol. 2015 May;9(3):625-31. doi: 10.1177/1932296815569246. Epub 2015 Jan 28.
9
Study on the dosing accuracy of commonly used disposable insulin pens.常用一次性胰岛素笔注射剂量准确性的研究。
Diabetes Technol Ther. 2012 Sep;14(9):804-9. doi: 10.1089/dia.2011.0298. Epub 2012 Jun 13.