瞳孔大小对使用IOLMaster进行生物测量的影响。

Effect of pupil size on biometry measurements using the IOLMaster.

作者信息

Adler Galia, Shahar Jonathan, Kesner Rivka, Rosenfeld Eldar, Fischer Naomi, Loewenstein Anat, Kurtz Shimon

机构信息

Hebrew University Hadassah Medical School, Jerusalem Israel.

Department of Ophthalmology, Tel Aviv Medical Center, Tel Aviv, Israel.

出版信息

Am J Ophthalmol. 2015 May;159(5):940-4. doi: 10.1016/j.ajo.2015.01.025. Epub 2015 Jan 28.

Abstract

PURPOSE

To evaluate the effect of pupil dilation on biometric measurements and intraocular lens power calculation using the IOLMaster (Carl Zeiss Meditec).

DESIGN

Prospective, observational case series.

METHODS

In this prospective study, 2 consecutive optical biometry measurements, before and after pupil dilation, were obtained using the IOLMaster on 318 eyes of 214 patients at the cataract presurgery clinic. The parameters compared were axial length, corneal power, cylinder, and the corresponding intraocular lens power, which was calculated using the Sanders-Retzlaff-Kraff/Theoretical formula.

RESULTS

This study found no statistically significant difference before and after dilation in axial length (0.005 mm; P = .476), corneal power (0.001 diopters [D]; P = .933), or calculated intraocular lens power (0.011 D; P = .609). A statistically significant difference was shown in cylinder measurements before and after dilation (0.102 D; P < .001).

CONCLUSIONS

This study demonstrated there is no clinically significant effect of pupil dilation on the IOLMaster measurements of axial length, corneal power, and corresponding theoretical intraocular lens power calculated using the Sanders-Retzlaff-Kraff/Theoretical formula.

摘要

目的

使用IOLMaster(卡尔蔡司医疗技术公司)评估瞳孔散大对生物测量及人工晶状体屈光力计算的影响。

设计

前瞻性观察性病例系列研究。

方法

在这项前瞻性研究中,在白内障术前门诊对214例患者的318只眼使用IOLMaster在瞳孔散大前后连续进行了两次光学生物测量。比较的参数包括眼轴长度、角膜屈光力、柱镜度数以及使用Sanders-Retzlaff-Kraff/理论公式计算的相应人工晶状体屈光力。

结果

本研究发现瞳孔散大前后眼轴长度(0.005 mm;P = 0.476)、角膜屈光力(0.001屈光度[D];P = 0.933)或计算的人工晶状体屈光力(0.011 D;P = 0.609)均无统计学上的显著差异。瞳孔散大前后柱镜测量结果显示有统计学上的显著差异(0.102 D;P < 0.001)。

结论

本研究表明,瞳孔散大对使用Sanders-Retzlaff-Kraff/理论公式计算的眼轴长度、角膜屈光力及相应理论人工晶状体屈光力的IOLMaster测量结果无临床显著影响。