The aim of this study was to evaluate the effect of pitch distance on the primary stability (PS) of orthodontic mini-implants (MIs) in artificial bone. Twenty experimental MIs were allocated to two groups, according to their geometric design: G1 (30o X 0.6 mm) and G2 (45o X 0.8 mm), and inserted into artificial bone of different densities: D1 (0.32 g/cm3) and D2 (0.64 g/cm3). The maximum insertion torque (IT) and removal torque (RT) values were recorded in N.cm. Loss of torque (LT) values were obtained by calculating the difference between the IT and RT. MI mobility was measured by means of a Periotest assessment. Normality and homogeneity were determined by means of the Kolmogorov-Smirnov and Shapiro-Wilk tests, respectively. A two-way ANOVA was used to detect differences between the mini-implant design and density factors. The ANOVA/Tukey tests were used to determine the intergroup difference. Higher IT values were observed for G2 (p ≤ 0.05) in D2. No statistical difference for RT was observed between the groups, whereas G2 presented higher values only for LT (p ≤ 0.05). The Periotest values (PTV) were higher for G1, in comparison with G2, in D1. G1 presented better PS in D2 (IT, RT and LT), whereas G2 was found to be more stable in D1, after evaluation with Periotest.
本研究的目的是评估螺距距离对正畸微型种植体(MIs)在人工骨中的初期稳定性(PS)的影响。根据几何设计,将20枚实验性微型种植体分为两组:G1(30°×0.6毫米)和G2(45°×0.8毫米),并植入不同密度的人工骨:D1(0.32克/立方厘米)和D2(0.64克/立方厘米)。以牛顿·厘米为单位记录最大插入扭矩(IT)和取出扭矩(RT)值。通过计算IT和RT之间的差值获得扭矩损失(LT)值。通过牙周探针评估测量微型种植体的松动度。分别通过柯尔莫哥洛夫-斯米尔诺夫检验和夏皮罗-威尔克检验确定正态性和同质性。采用双向方差分析来检测微型种植体设计和密度因素之间的差异。采用方差分析/图基检验来确定组间差异。在D2中,观察到G2的IT值更高(p≤0.05)。两组之间未观察到RT的统计学差异,而G2仅在LT方面呈现更高的值(p≤0.05)。在D1中,与G2相比,G1的牙周探针值(PTV)更高。在用牙周探针评估后,G1在D2中的PS更好(IT、RT和LT),而G2在D1中更稳定。