角膜地形图散光(CorT)用于量化总角膜散光。

Corneal topographic astigmatism (CorT) to quantify total corneal astigmatism.

作者信息

Alpins Noel, Ong James K Y, Stamatelatos George

出版信息

J Refract Surg. 2015 Mar;31(3):182-6. doi: 10.3928/1081597X-20150224-02.

Abstract

PURPOSE

To evaluate the performance of corneal topographic astigmatism (CorT) based on total corneal power measurements.

METHODS

Anterior, posterior, and total corneal power measurements of 526 virgin eyes obtained using the CSO Sirius tomographer (Costuzione Strumenti Oftalmici, Scandicci, Florence, Italy) were analyzed. Individual CorTs were created from each set of data. These CorTs were assessed using ocular residual astigmatism (ORA), which quantifies corneo-refractive differences. A low standard deviation of the ocular residual astigmatism (ORAsd) indicates a low variability between corneal astigmatism and refractive cylinder. A low mean of the ORA magnitude indicates a close correlation of refractive cylinder and corneal astigmatism.

RESULTS

The CorT based on total corneal power measurements had an ORAsd of 0.30 diopters (D) and a mean ORA magnitude of 0.53 D. The CorT candidates based on anterior corneal power measurements all had an ORAsd of at least 0.32 D, and the mean ORA magnitudes were all 0.64 D or greater. Both the ORAsd and mean ORA magnitude of the CorT based on total corneal power measurements were significantly less than those of the CorT based on anterior corneal power measurements (both P < .001, as estimated via bootstrapping).

CONCLUSIONS

The CorT based on total corneal power measurements corresponds better, both in variability and closeness, with manifest refractive cylinder than the CorT based on anterior corneal power measurements. This total CorT would be fundamental when planning toric intraocular lenses or limbal relaxing incisions or other corneal astigmatic surgery.

摘要

目的

基于全角膜屈光力测量评估角膜地形图散光(CorT)的性能。

方法

分析使用CSO Sirius眼前节分析仪(意大利佛罗伦萨斯坎迪奇的Costuzione Strumenti Oftalmici公司生产)获得的526只未手术眼的前表面、后表面和全角膜屈光力测量值。从每组数据中创建个体CorT。使用眼残余散光(ORA)评估这些CorT,ORA可量化角膜屈光差异。眼残余散光标准差(ORAsd)低表明角膜散光和屈光柱镜之间的变异性低。ORA大小的平均值低表明屈光柱镜和角膜散光之间密切相关。

结果

基于全角膜屈光力测量的CorT的ORAsd为0.30屈光度(D),ORA大小的平均值为0.53 D。基于前表面角膜屈光力测量的CorT候选值的ORAsd均至少为0.32 D,ORA大小的平均值均为0.64 D或更高。基于全角膜屈光力测量的CorT的ORAsd和ORA大小平均值均显著低于基于前表面角膜屈光力测量的CorT(通过自助法估计,两者P <.001)。

结论

与基于前表面角膜屈光力测量的CorT相比,基于全角膜屈光力测量的CorT在变异性和接近程度方面与明显的屈光柱镜更相符。在规划散光人工晶状体、角膜缘松解切口或其他角膜散光手术时,这种全角膜CorT将至关重要。