基于总角膜力数据的角膜地形散光(CorT Total):总角膜散光的基准。

Corneal Topographic Astigmatism Based on Total Corneal Power Data (CorT Total): A Benchmark for Total Corneal Astigmatism.

机构信息

NewVision Clinics; Melbourne, Australia.

Department of Ophthalmology, University of Melbourne, Melbourne, Australia.

出版信息

Cornea. 2020 Apr;39(4):431-436. doi: 10.1097/ICO.0000000000002162.

Abstract

PURPOSE

To evaluate how closely manufacturer-provided measures of total corneal astigmatism correspond with the manifest refractive cylinder, as compared to a benchmark of corneal topographic astigmatism calculated on the basis of measured total corneal power (TCP) data (CorT Total).

METHODS

The SD of the ocular residual astigmatism magnitude (ORAsd) was evaluated for normal virgin eyes based on an optimized benchmark CorT Total and the various measures of total corneal astigmatism provided by 3 different Scheimpflug tomographers.

RESULTS

The CorT Total corresponded with the manifest refractive cylinder at least as well as all the measures of total corneal astigmatism provided by the tomographers [Sirius CorT Total ORAsd: 0.320D (standard error [SE] 0.017D), Sirius TCP 4 mm ORAsd: 0.324D (SE 0.017D); Pentacam CorT Total ORAsd: 0.338D (SE 0.027D), Pentacam total corneal refractive power apex zone 4 mm ORAsd: 0.337D (SE 0.029D); Galilei CorT Total ORAsd: 0.472D (SE 0.068D), and Galilei TCP2 ORAsd: 0.536D (SE 0.124D)]. The difference between CorT Total and best measure on each tomographer was not statistically significant (Sirius TCP 4 mm: P = 0.24, Pentacam total corneal refractive power apex zone 4 mm: P = 0.64, Galilei TCP2: P = 0.24). Most of the manufacturer-provided measures did not correspond closely with the manifest refractive cylinder. When there were multiple measures of total corneal astigmatism, those derived from a zone with a diameter of 4.0 mm corresponded best with the manifest refractive cylinder.

CONCLUSIONS

The CorT Total is a reliable benchmark measure that can be used to assess how well other measures of total corneal astigmatism correspond with the manifest refractive cylinder.

摘要

目的

评估制造商提供的总角膜散光测量值与基于实测角膜总屈光力 (TCP) 数据(CorT Total)计算的角膜地形图散光基准值(CorT Total)之间的吻合程度,与基于实测角膜总屈光力(TCP)数据(CorT Total)计算的角膜地形图散光基准值(CorT Total)相比。

方法

基于优化的 CorT Total 和三种不同的Scheimpflug 断层扫描仪提供的总角膜散光的各种测量值,评估正常初眼的眼残余散光幅度(ORAsd)的标准差。

结果

CorT Total 与角膜曲率计测量的等效球镜值至少一样吻合,优于所有断层扫描仪提供的总角膜散光的测量值(Sirius CorT Total ORAsd:0.320D(标准误差 [SE] 0.017D), Sirius TCP 4mm ORAsd:0.324D(SE 0.017D);Pentacam CorT Total ORAsd:0.338D(SE 0.027D),Pentacam 角膜总屈光力顶点区 4mm ORAsd:0.337D(SE 0.029D);Galilei CorT Total ORAsd:0.472D(SE 0.068D),Galilei TCP2 ORAsd:0.536D(SE 0.124D))。CorT Total 与每个断层扫描仪上最佳测量值之间的差异无统计学意义(Sirius TCP 4mm:P=0.24,Pentacam 角膜总屈光力顶点区 4mm:P=0.64,Galilei TCP2:P=0.24)。制造商提供的大多数总角膜散光的测量值与角膜曲率计测量的等效球镜值吻合不佳。当有多个总角膜散光的测量值时,直径为 4.0mm 的区域内的测量值与角膜曲率计测量的等效球镜值最吻合。

结论

CorT Total 是一种可靠的基准测量值,可用于评估其他总角膜散光的测量值与角膜曲率计测量的等效球镜值的吻合程度。