Suppr超能文献

一种用于测定流式细胞仪扩增系统线性的简单快速方法。

A simple and rapid method for determining the linearity of a flow cytometer amplification system.

作者信息

Bagwell C B, Baker D, Whetstone S, Munson M, Hitchcox S, Ault K A, Lovett E J

机构信息

Maine Cytometry Research Institute, Portland 04102.

出版信息

Cytometry. 1989 Nov;10(6):689-94. doi: 10.1002/cyto.990100604.

Abstract

We describe a simple and rapid method for determining the linearity of a flow cytometer amplification system. The method is based on a fundamental characteristic of linear amplifiers: The difference between two amplified signals increases linearly with increasing amplifier gain. Two populations of beads or cells, differing slightly in fluorescence intensity, are analyzed by the flow cytometer at increasing photomultiplier tube high-voltage settings. The distribution of the populations' mean difference versus mean position is a straight line intersecting the origin for linear amplifiers. Although some types of nonlinearities cannot be detected with this technique, deviations from linearity indicate nonlinear components in the flow cytometer amplification system. The correlation coefficient is used to quantify degree of nonlinearity. We also describe a method for amplifier nonlinearity compensation.

摘要

我们描述了一种用于确定流式细胞仪放大系统线性的简单快速方法。该方法基于线性放大器的一个基本特性:两个放大信号之间的差异随放大器增益的增加而线性增加。在光电倍增管高压设置不断增加的情况下,流式细胞仪对荧光强度略有不同的两组微珠或细胞进行分析。对于线性放大器,两组的平均差异与平均位置的分布是一条与原点相交的直线。尽管该技术无法检测到某些类型的非线性,但与线性的偏差表明流式细胞仪放大系统中存在非线性组件。相关系数用于量化非线性程度。我们还描述了一种放大器非线性补偿方法。