C型II区骨盆骨折模型中三种内固定方式的生物力学比较

Biomechanical comparison of three types of internal fixation in a type C zone II pelvic fracture model.

作者信息

Wu Tao, Chen Wei, Li Xu, Zhang Qi, Lv Hong-Zhi, Zhang Ying-Ze

机构信息

Emergency Center of Trauma, Key Laboratory of Orthopaedic Biomechanics of Hebei Province, The Third Hospital of Hebei Medical University, Orthopaedic Research Institution of Hebei Province Shijiazhuang 050051, China.

出版信息

Int J Clin Exp Med. 2015 Feb 15;8(2):1853-61. eCollection 2015.

Abstract

OBJECTIVE

This study aimed to compare the stability of Tile C pelvic fractures fixed with two iliosacral (IS) screws, tension band plate (TBP), and minimally invasive adjustable plate (MIAP).

METHODS

Six embalmed specimens of adult pelvis were used. The soft tissue was removed from the specimens, but spines from the fourth lumbar vertebra to the proximal one-third of both femurs were retained. The pubic symphysis, bilateral sacroiliac joints and ligaments, bilateral hip joints, bilateral sacrotuberous ligaments, and bilateral sacrospinous ligaments were intact. Tile C pelvic fractures were created on the specimens. The symphysis pubis was fixed with a plate, and the fracture on the posterior pelvic ring was fixed with three types of internal fixation in a randomized block design. The specimens were placed in a biomechanical machine at a standing neutral posture. A cyclic vertical load of up to 500N was applied, and displacement was recorded. Shifts in the fracture gap were measured by a grating displacement sensor.

RESULTS

Under different vertical loads, the shift in the fracture gap and displacement of the pelvic fractures fixed with two IS screws were similar to those in fractures fixed with MIAP. However, the shift in the fracture gap and displacement of fractures fixed with MIAP was significantly smaller than those of fractures fixed with TBP.

CONCLUSION

The stability of the Tile C pelvic fractures fixed with MIAP was similar to that of fractures fixed with IS screws. MIAP performed better than TBP under vertical load.

摘要

目的

本研究旨在比较使用两枚髂骶(IS)螺钉、张力带钢板(TBP)和微创可调钢板(MIAP)固定Tile C型骨盆骨折的稳定性。

方法

使用六个成年骨盆防腐标本。从标本上移除软组织,但保留从第四腰椎到双侧股骨近端三分之一的棘突。耻骨联合、双侧骶髂关节及韧带、双侧髋关节、双侧骶结节韧带和双侧骶棘韧带均保持完整。在标本上制造Tile C型骨盆骨折。耻骨联合用钢板固定,骨盆后环骨折采用随机区组设计用三种内固定方式固定。将标本以站立中立姿势放置在生物力学机器中。施加高达500N的循环垂直载荷,并记录位移。用光栅位移传感器测量骨折间隙的移位。

结果

在不同垂直载荷下,用两枚IS螺钉固定的骨盆骨折的骨折间隙移位和位移与用MIAP固定的骨折相似。然而,用MIAP固定的骨折的骨折间隙移位和位移明显小于用TBP固定的骨折。

结论

用MIAP固定的Tile C型骨盆骨折的稳定性与用IS螺钉固定的骨折相似。在垂直载荷下,MIAP的性能优于TBP。

相似文献

引用本文的文献

本文引用的文献

9
Haemodynamically unstable pelvic fractures.血流动力学不稳定的骨盆骨折。
Injury. 2009 Oct;40(10):1023-30. doi: 10.1016/j.injury.2008.11.023. Epub 2009 Apr 16.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验