三种pH值及血气分析仪电极系统的比较研究

A comparative study of the electrode systems of three pH and blood gas apparatus.

作者信息

Winckers E K, Teunissen A J, Van den Camp R A, Maas A H

出版信息

J Clin Chem Clin Biochem. 1978 Mar;16(3):175-85. doi: 10.1515/cclm.1978.16.3.175.

Abstract

We present a comparative evaluation of the electrode systems of three modern blood gas analysers: IL-413, ABL-1 and AVL-937C. The response curves, accuracy and precision of the pH-, pCO2- and pO2-electrodes were established with tonometered blood and buffer solutions. pH values (range 6.8-7.8) measured on the AVL deviate (-0.03 pH for blood and +0.03 pH for buffer) from those of BMS2 Mk2; whereas on the IL and ABL analysers the pH values deviate by not more than 0.01 pH. The standard deviation was better than 0.005 pH. pCO2 values of blood and buffer (range 14-106 mm Hg) deviate from the calculated tonometer values by quantities ranging from 3 to 10 mm Hg. The average precision (CV)1) of the pCO2 measurement on each analyser was better than 1.8%. pO2 values of blood (range 0-130 mm Hg) did not differ by more than 3 mm Hg from the calculated values. Above 130 mm Hg a linear negative increasing difference was seen. For buffer solutions a linear relationship between pO2 difference and pO2 value was found over the whole range from zero up to 642 mm Hg: a positive difference below and a negative difference above the pO2 of the previous calibration; if the calibration pO2 is higher, the sample pO2 is shifted to a higher value. The average precision of the pO2 measurements was better than 3%. In the (patho)-physiological range the three instruments may provide suitable results for the clinician. Suggestions are made for standardization and improvement of the electrode systems.

摘要

我们对三款现代血气分析仪(IL-413、ABL-1和AVL-937C)的电极系统进行了比较评估。使用经校准的血液和缓冲溶液确定了pH电极、pCO₂电极和pO₂电极的响应曲线、准确性和精密度。在AVL上测得的pH值(范围为6.8 - 7.8)与BMS2 Mk2的pH值存在偏差(血液偏差-0.03 pH,缓冲溶液偏差+0.03 pH);而在IL和ABL分析仪上,pH值偏差不超过0.01 pH。标准偏差优于0.005 pH。血液和缓冲溶液的pCO₂值(范围为14 - 106 mmHg)与计算得出的校准值偏差为3至10 mmHg。各分析仪上pCO₂测量的平均精密度(CV)1)优于1.8%。血液的pO₂值(范围为0 - 130 mmHg)与计算值的差异不超过3 mmHg。高于130 mmHg时,出现线性负向增加的差异。对于缓冲溶液,在从零到642 mmHg的整个范围内,发现pO₂差值与pO₂值之间存在线性关系:在校准pO₂以下为正差值,在校准pO₂以上为负差值;如果校准pO₂较高,则样品pO₂会向更高值偏移。pO₂测量的平均精密度优于3%。在(病理)生理范围内,这三款仪器可为临床医生提供合适的结果。针对电极系统的标准化和改进提出了建议。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验