Plaza-Puche Ana B, Alió Jorge L, MacRae Scott, Zheleznyak Len, Sala Esperanza, Yoon Geunyoung
J Refract Surg. 2015 May;31(5):300-7. doi: 10.3928/1081597X-20150423-03.
To investigate the correlations existing between a trifocal intraocular lens (IOL) and a varifocal IOL using the "ex vivo" optical bench through-focus image quality analysis and the clinical visual performance in real patients by study of the defocus curves.
This prospective, consecutive, nonrandomized, comparative study included a total of 64 eyes of 42 patients. Three groups of eyes were differentiated according to the IOL implanted: 22 eyes implanted with the varifocal Lentis Mplus LS-313 IOL (Oculentis GmbH, Berlin, Germany); 22 eyes implanted with the trifocal FineVision IOL (Physiol, Liege, Belgium), and 20 eyes implanted with the monofocal Acrysof SA60AT IOL (Alcon Laboratories, Inc., Fort Worth, TX). Visual outcomes and defocus curve were evaluated postoperatively. Optical bench through-focus performance was quantified by computing an image quality metric and the cross-correlation coefficient between an unaberrated reference image and captured retinal images from a model eye with a 3.0-mm artificial pupil.
Statistically significant differences among defocus curves of different IOLs were detected for the levels of defocus from -4.00 to -1.00 diopters (D) (P < .01). Significant correlations were found between the optical bench image quality metric results and logMAR visual acuity scale in all groups (Lentis Mplus group: r = -0.97, P < .01; FineVision group: r = -0.82, P < .01; Acrys of group: r = -0.99, P < .01). Linear predicting models were obtained.
Significant correlations were found between logMAR visual acuity and image quality metric for the multifocal and monofocal IOLs analyzed. This finding enables surgeons to predict visual outcomes from the optical bench analysis.
通过“离体”光学平台的全聚焦图像质量分析以及对真实患者的散焦曲线研究,探讨三焦点人工晶状体(IOL)和变焦人工晶状体之间存在的相关性。
这项前瞻性、连续性、非随机对照研究共纳入42例患者的64只眼。根据植入的人工晶状体将眼部分为三组:22只眼植入变焦Lentis Mplus LS - 313人工晶状体(德国柏林Oculentis GmbH公司);22只眼植入三焦点FineVision人工晶状体(比利时列日Physiol公司);20只眼植入单焦点Acrysof SA60AT人工晶状体(美国德克萨斯州沃思堡爱尔康实验室公司)。术后评估视觉结果和散焦曲线。通过计算图像质量指标以及具有3.0毫米人工瞳孔的模型眼中未矫正参考图像与捕获的视网膜图像之间的互相关系数,对光学平台全聚焦性能进行量化。
在-4.00至-1.00屈光度(D)的散焦水平下,检测到不同人工晶状体的散焦曲线之间存在统计学显著差异(P <.01)。在所有组中,光学平台图像质量指标结果与logMAR视力表之间均发现显著相关性(Lentis Mplus组:r = -0.97,P <.01;FineVision组:r = -0.82,P <.01;Acrysof组:r = -0.99,P <.01)。获得了线性预测模型。
在所分析的多焦点和单焦点人工晶状体中,logMAR视力与图像质量指标之间存在显著相关性。这一发现使外科医生能够通过光学平台分析预测视觉结果。