Repeatability data and agreement of keratometry with the VERION system compared to the IOLMaster.

作者信息

Nemeth Gabor, Szalai Eszter, Hassan Ziad, Lipecz Agnes, Berta Andras, Modis Laszlo

出版信息

J Refract Surg. 2015 May;31(5):333-7. doi: 10.3928/1081597X-20150424-01.

Abstract

PURPOSE

To analyze the repeatability of keratometric and white-to-white (WTW) distance measurements with the VERION Measurement Module (Alcon Laboratories, Inc., Fort Worth, TX) and to compare the measured data to the results of the IOLMaster (Carl Zeiss Meditec, Jena, Germany).

METHODS

Three images were captured with the VERION and the flattest and steepest keratometric data, the astigmatism axis, and the WTW distance were recorded. Subsequently, the axial length, the keratometric data with axis, and the WTW distance were measured with an IOLMaster. The repeatability data of the keratometric value of the VERION System, converted to cross cylinder J0 and J45 vector components, were analyzed. The agreement data for keratometry obtained by the VERION System and the differences regarding keratometric data and WTW distance compared to IOLMaster were calculated.

RESULTS

The measurements were conducted in 50 eyes of 50 healthy volunteers (median age: 50.32 years, range: 19.34 to 85.3 years). The flattest and the steepest keratometric data, the diopter of astigmatism, the J0 and J45 vector components, and WTW distance did not differ significantly between devices (P > .05). Intraclass correlation coefficients (range: 0.863 to 0.994) and Cronbach's alpha values (range: 0.950 to 0.998) were high for all parameters measured by the VERION System.

CONCLUSIONS

The VERION System has high repeatability and agreement with the IOLMaster, making it suitable as an alternative tool in clinical practice.

摘要

目的

分析使用VERION测量模块(爱尔康实验室公司,沃思堡,德克萨斯州)进行角膜曲率和白到白(WTW)距离测量的可重复性,并将测量数据与IOLMaster(卡尔蔡司医疗技术公司,耶拿,德国)的结果进行比较。

方法

使用VERION拍摄三张图像,并记录最平坦和最陡峭的角膜曲率数据、散光轴和WTW距离。随后,使用IOLMaster测量眼轴长度、带轴的角膜曲率数据和WTW距离。分析了VERION系统角膜曲率值的可重复性数据,将其转换为交叉柱镜J0和J45矢量分量。计算了VERION系统获得的角膜曲率测量的一致性数据,以及与IOLMaster相比角膜曲率数据和WTW距离的差异。

结果

对50名健康志愿者的50只眼睛进行了测量(年龄中位数:50.32岁,范围:19.34至85.3岁)。各设备之间最平坦和最陡峭的角膜曲率数据、散光度数、J0和J45矢量分量以及WTW距离无显著差异(P>.05)。VERION系统测量的所有参数的组内相关系数(范围:0.863至0.994)和克朗巴哈系数(范围:0.950至0.998)都很高。

结论

VERION系统具有较高的可重复性,并且与IOLMaster一致性良好,使其适合作为临床实践中的替代工具。