三种不同设备所提供的角膜曲率和非球面性测量结果的互换性。

Interchangeability of corneal curvature and asphericity measurements provided by three different devices.

作者信息

Piñero David P, Soto-Negro Roberto, Ruiz-Fortes Pedro, Pérez-Cambrodí Rafael J, Fukumitsu Hideki

机构信息

Department of Optics, Pharmacology and Anatomy, University of Alicante, San Vicente del Raspeig 03690, Alicante, Spain.

Department of Ophthalmology (OFTALMAR), Vithas Medimar International Hospital, Alicante 03014, Spain.

出版信息

Int J Ophthalmol. 2019 Mar 18;12(3):412-416. doi: 10.18240/ijo.2019.03.10. eCollection 2019.

Abstract

AIM

To evaluate the interchangeability of keratometric and asphericity measurements provided by three measurement systems based on different optical principles.

METHODS

A total of 40 eyes of 40 patients with a mean age of 34.1y were included. In all cases, a corneal curvature analysis was performed with IOL-Master (IOLM), iDesign 2 (ID2), and Sirius systems (SIR). Differences between instruments for flattest (K1) and steepest (K2) keratometric readings, as well as for magnitude and axis of corneal astigmatism were analyzed. Likewise, differences in asphericity (Q) between SIR and ID2 were also evaluated.

RESULTS

Mean differences between devices for K1 were 0.20±0.21 (<0.001), -0.12±0.36 (=0.046) and -0.32±0.36 D (<0.001) for the comparisons IOLM-SIR, IOLM-ID2 and SIR-ID2, respectively. The ranges of agreement for these comparisons between instruments were 0.41, 0.70, and 0.70 D. For K2, mean differences were 0.31±0.33 (<0.001), -0.08±0.43 (=0.265) and -0.39±0.38 D (<0.001), with ranges of agreement of 0.65, 0.84, and 0.74 D. Concerning magnitude of astigmatism, ranges of agreement were in the limit of clinical relevance (0.49 D, =0.011; 0.55 D, =0.386; 0.43 D, =0.05). In contrast, ranges of agreement were clinically relevant for astigmatic axis (26.68°, 33.83° and 18.37°, ≥0.121) and for Q between SIR and ID2 (0.16, <0.001).

CONCLUSION

The keratometric corneal power, astigmatic axis and asphericity measurements provide by the three systems evaluated cannot be considered as interchangeable, whereas measurements of corneal astigmatism obtained with SIR and ID2 can be considered as interchangeable for clinical purposes.

摘要

目的

评估基于不同光学原理的三种测量系统所提供的角膜曲率和非球面性测量值的互换性。

方法

纳入40例患者的40只眼,平均年龄34.1岁。所有病例均使用IOL-Master(IOLM)、iDesign 2(ID2)和Sirius系统(SIR)进行角膜曲率分析。分析各仪器在最平(K1)和最陡(K2)角膜曲率读数以及角膜散光度数和轴向上的差异。同样,也评估了SIR和ID2之间非球面性(Q)的差异。

结果

IOLM-SIR、IOLM-ID2和SIR-ID2比较中,K1测量值的仪器间平均差异分别为0.20±0.21(<0.001)、-0.12±0.36(=0.046)和-0.32±0.36 D(<0.001)。这些仪器间比较的一致性范围分别为0.41、0.70和0.70 D。对于K2,平均差异分别为0.31±0.33(<0.001)、-0.08±0.43(=0.265)和-0.39±0.38 D(<0.001),一致性范围分别为0.65、0.84和0.74 D。关于散光度数,一致性范围处于临床相关限度(0.49 D,=0.011;0.55 D,=至0.386;0.43 D,=0.05)。相比之下,散光轴(26.68°、33.83°和18.37°,≥0.121)以及SIR和ID2之间的Q(0.16,<0.001)的一致性范围具有临床相关性。

结论

所评估的三种系统提供的角膜曲率、散光轴和非球面性测量值不能被视为可互换的,而SIR和ID2获得的角膜散光测量值在临床应用中可被视为可互换的。