Suppr超能文献

Verion图像引导系统与自动验光角膜曲率计所获得的生物测量结果比较。

Comparison of biometric measurements obtained by the Verion Image-Guided System versus the auto-refracto-keratometer.

作者信息

Velasco-Barona Cecilio, Cervantes-Coste Guadalupe, Mendoza-Schuster Erick, Corredor-Ortega Claudia, Casillas-Chavarín Nadia L, Silva-Moreno Alejandro, Garza-León Manuel, Gonzalez-Salinas Roberto

机构信息

Anterior Segment Surgery Department, Asociación para Evitar la Ceguera en México, 04030, Mexico City, Mexico.

Dirección de Ciencias Médicas de la División de Ciencias de la Salud, Universidad de Monterrey, Monterrey, Nuevo Leon, Mexico.

出版信息

Int Ophthalmol. 2018 Jun;38(3):951-957. doi: 10.1007/s10792-017-0541-3. Epub 2017 Apr 25.

Abstract

PURPOSE

To compare the biometric measurements obtained from the Verion Image-Guided System to those obtained by auto-refracto-keratometer in normal eyes.

METHODS

This is a prospective, observational, comparative study conducted at the Asociación para Evitar la Ceguera en México I.A.P., Mexico. Three sets of keratometry measurements were obtained using the image-guided system to assess the coefficient of variation, the within-subject standard deviation and intraclass correlation coefficient (ICC). A paired Student t test was used to assess statistical significance between the Verion and the auto-refracto-keratometer. A Pearson's correlation coefficient (r) was obtained for all measurements, and the level of agreement was verified using Bland-Altman plots.

RESULTS

The right eyes of 73 patients were evaluated by each platform. The Verion coefficient of variation was 0.3% for the flat and steep keratometry, with the ICC being greater than 0.9 for all parameters measured. Paired t test showed statistically significant differences between groups (P = 0.0001). A good correlation was evidenced for keratometry values between platforms (r = 0.903, P = 0.0001 for K1, and r = 0.890, P = 0.0001). Bland-Altman plots showed a wide data spread for all variables.

CONCLUSION

The image-guided system provided highly repeatable corneal power and keratometry measurements. However, significant differences were evidenced between the two platforms, and although values were highly correlated, they showed a wide data spread for all analysed variables; therefore, their interchangeable use for biometry assessment is not advisable.

摘要

目的

比较在正常眼中使用Verion图像引导系统获得的生物测量数据与自动验光角膜曲率计获得的数据。

方法

这是一项在墨西哥墨西哥预防失明协会进行的前瞻性、观察性、对比研究。使用图像引导系统获得三组角膜曲率测量数据,以评估变异系数、受试者内标准差和组内相关系数(ICC)。采用配对学生t检验评估Verion系统与自动验光角膜曲率计之间的统计学显著性。对所有测量数据计算Pearson相关系数(r),并使用Bland-Altman图验证一致性水平。

结果

每个平台评估了73例患者的右眼。Verion系统平角膜曲率和陡角膜曲率的变异系数均为0.3%,所有测量参数的ICC均大于0.9。配对t检验显示两组之间存在统计学显著差异(P = 0.0001)。两个平台之间的角膜曲率值具有良好的相关性(K1的r = 0.903,P = 0.0001;K2的r = 0.890,P = 0.0001)。Bland-Altman图显示所有变量的数据分布较广。

结论

图像引导系统提供了高度可重复的角膜屈光力和角膜曲率测量。然而,两个平台之间存在显著差异,尽管数值高度相关,但所有分析变量的数据分布较广;因此,不建议将它们互换用于生物测量评估。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验